PUMP MMT-1780KPK 670G PATHWAY BLACK MG
Report
- Report Number
- 2032227-2021-181128
- Event Type
- Malfunction
- Date Received
- August 17, 2021
- Date of Event
- August 6, 2021
- Report Date
- November 12, 2021
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OZP
- UDI-DI
- 000000763000393786
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- OTHER
- Health Professional
- N
Narratives
ON (B)(6) 2021 THE CUSTOMER ALLEGED THE INSULIN PUMP ALARMED CRITICAL PUMP ERROR/OPEN BOOK IMAGE. DEVICE RECEIVED WITH CRITICAL PUMP ERROR (OPEN BOOK) AND CREST DOWNLOAD UNSUCCESSFUL DUE TO MOISTURE DAMAGE ON ELECTRONICS ASSEMBLY STACK ELECTRICAL BOARD 1 AND ELECTRICAL BOARD 2. UNABLE TO PERFORMED DISPLACEMENT TEST, REWIND TEST, PRIME/SEATING TEST, BASIC OCCLUSION TEST, FORCE SENSOR TEST, OCCLUSION TEST, SLEEP CURRENT MEASUREMENT, ACTIVE CURRENT MEASUREMENT AND SELF TEST DUE TO CRITICAL PUMP ERROR. DEVICE RECEIVED WITH MOISTURE DAMAGE NOTED ON, 40 PIN FLEXIBLE PRINTED CIRCUIT/LCD, ELECTRONICS STACK ELECTRICAL BOARD 1, ELECTRICAL BOARD 2, MOTOR AND BATTERY TUBE. HOWEVER, DEVICE TESTED WITH REFERENCE TEST ELECTRONICS STACK ELECTRICAL BOARD 2, WAS ABLE TO BOOT UP PROPERLY WITH NO UNEXPECTED CRITICAL PUMP ERROR ALARM NOTED. DEVICE RECEIVED WITH CRACKED BATTERY TUBE THREADS, AND CRACKED CASE CORNER OF BELT CLIP RAILS. THE DEVICE P-CAP / TEST RESERVOIR LOCKS IN PLACE PROPERLY. IN SUMMARY, CUSTOMER ALLEGATION FOR CRITICAL PUMP ERROR ALARM WAS CONFIRMED, HOWEVER, DEVICE TESTED WITH REFERENCE TEST ELECTRONICS STACK ELECTRICAL BOARD 2, WAS ABLE TO BOOT UP PROPERLY WITH NO UNEXPECTED CRITICAL PUMP ERROR ALARM NOTED. MEDTRONIC, INC. (MEDTRONIC) IS SUBMITTING THIS REPORT TO COMPLY WITH 21 C.F.R. PART 803, THE MEDICAL DEVICE REPORTING REGULATION. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION IN THE TIME ALLOTTED AND HAS PROVIDED AS MUCH INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY "DEFECTS" OR HAS "MALFUNCTIONED". THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA, TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REPORTING PURSUANT TO PART 803. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE IT BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT.
(B)(4). CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD OTHER CRITICAL PUMP ERROR ALARM. CUSTOMER WENT SWIMMING WITH INSULIN PUMP. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1231719 | PUMP MMT-1780KPK 670G PATHWAY BLACK MG | ARTIFICIAL PANCREAS DEVICE SYSTEM, SINGLE HORMONAL CONTROL | OZP | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-1780KPK | HG5884G | 000000763000393786 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 54 YR | Female |