FDA Adverse Event
Injury
Summary report: N
GENTLEPOWER LUX CONTRA ANGLE 25LPA
MDR report key: 1232923
·
Received November 18, 2008
Report
- Report Number
- 8010493-2008-00023
- Event Type
- Injury
- Date Received
- November 18, 2008
- Date of Event
- November 5, 2008
- Report Date
- November 7, 2008
- Manufacturer
- KALTENBACH & VOIGT GMBH
- Product Code
- EFB
- PMA / PMN Number
- K073478
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- KY, US
- Reporter Occupation
- DENTIST
Narratives
Additional Manufacturer Narrative · 1
THE PATIENT WAS PRESCRIBED KENALOG OINTMENT FOR TREATMENT OF BURN TO LIP. AT THE HANDPIECE EVALUATION, IT WAS FOUND THAT THE BEARINGS WERE WORN AND GRITTY IN DRIVE ASSEMBLY, SHAFT, AND AXLE, AND THE WATER PORT WAS CLOGGED. MEDIUM DEBRIS LEVEL WAS PRESENT. PROPER MAINTENANCE WAS DISCUSSED WITH THE OFFICE AND A TWO-YEAR SERVICE CHECK WAS SUGGESTED. THE HANDPIECE WAS REPAIRED.
Description of Event or Problem · 1
WHILE THE DENTIST WAS WORKING ON TOOTH #1, #2 AND #30, THE PATIENT'S LOWER LIP WAS BURNED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | GENTLEPOWER LUX CONTRA ANGLE 25LPA | DENTAL HANDPIECE AND ACCESSORIES | EFB | KALTENBACH & VOIGT GMBH | 25LPA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |