BD BBL¿ TRYPTICASE¿ SOY AGAR (TSA II) // MACCONKEY II AGAR
Report
- Report Number
- 1119779-2021-01393
- Event Type
- Malfunction
- Date Received
- August 16, 2021
- Date of Event
- August 4, 2021
- Report Date
- April 7, 2022
- Manufacturer
- BECTON, DICKINSON & CO. (SPARKS)
- Product Code
- JSG
- UDI-DI
- 10382902212918
- PMA / PMN Number
- PREAMENDMENT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- BE
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
H.6. INVESTIGATION: DURING MANUFACTURING OF MATERIAL 221291, MEDIA IS FORMULATED AND SENT THROUGH A HIGH TEMPERATURE SHORT TIME STERILIZER TO REMOVE BIOBURDEN. THE PETRI DISHES ARE SUBJECTED TO UV RADIATION TO DECREASE BIOBURDEN. THE PETRI DISHES ARE FILLED IN A POSITIVE PRESSURE HEPA FILTERED ENVIRONMENT. THE FILLED PLATES ARE COOLED AND IMMEDIATELY WRAPPED INTO SLEEVES TO DECREASE THE INTRODUCTION OF MICROBES. SLEEVES ARE THEN PACKAGED INTO CARTONS AND THEN TRANSFERRED TO A REFRIGERATED TRUCK (2 TO 8 DEGREES C) FOR SHIPMENT TO THE DISTRIBUTOR. BD DISTRIBUTORS ARE PROVIDED WITH THE STORAGE GUIDELINES FOR THE SHIPPING AND HANDLING OF BD MEDIA OF 2 TO 8 DEGREES C IN A DARK PLACE. THE BATCH HISTORY RECORD FOR BATCH 1120379 WAS SATISFACTORY AND NO QUALITY NOTIFICATIONS WERE GENERATED DURING MANUFACTURING AND INSPECTION. THE RELEASE TESTING THAT IS PERFORMED ON THIS PRODUCT DOES INCLUDE BIOBURDEN TESTING. A SAMPLE OF PLATES ARE INCUBATED AT 25 DEGREES C AND AT 35 DEGREES C FOR APPROXIMATELY 72 HOURS. ALL BIOBURDEN TESTING PERFORMED ON THIS BATCH WAS SATISFACTORY PER BD INTERNAL PROCEDURES. AFFECTED PRODUCT DOES NOT HAVE ANY STERILITY CLAIMS; THE PRODUCT IS TESTED FOR BIOBURDEN PRIOR TO RELEASE TO ENSURE THAT IT CONFORMS TO PRODUCT SPECIFICATIONS. HOWEVER, THIS DOES NOT ENSURE THAT THE END-USER WILL NOT RECEIVE A CONTAMINATED PLATE. THE COMPLAINT HISTORY WAS REVIEWED, AND NO OTHER COMPLAINTS HAVE BEEN TAKEN ON BATCH 1120379. RETENTION SAMPLES FROM BATCH 1120379 WERE NOT AVAILABLE FOR INSPECTION. TWO PHOTOS WERE RECEIVED FOR INVESTIGATION. ONE PHOTO SHOWS THE AGAR SURFACE OF EIGHT, OPENED PLATES FROM BATCH 1120379 (TIME STAMPS NOT READABLE) WITH SURFACE BACTERIAL GROWTH IN AT LEAST ONE OF THE MEDIA. THE OTHER PHOTO SHOWS THE BOTTOM OF A PLATE FROM BATCH 1120379 (TIME STAMP 0350) WITH BACTERIAL GROWTH IN BOTH MEDIA. NO RETURN SAMPLES WERE RECEIVED FOR INVESTIGATION. THIS COMPLAINT HAS BEEN CONFIRMED. BD WILL CONTINUE TO TREND COMPLAINTS FOR CONTAMINATION. A TREND WAS IDENTIFIED FOR CONTAMINATION AND INVESTIGATION FOUND OPPORTUNITIES FOR BIOBURDEN REDUCTION IN THE MANUFACTURING PROCESS. CAPA#3076308 HAS BEEN INITIATED AND INVOLVES IMPLEMENTING ADDITIONAL CLEANING EVENTS AND EVALUATION OF MANUFACTURING PROCEDURES FOCUSED ON IN-PROCESS BIOBURDEN REDUCTION. ADDITIONAL TRAININGS ARE PLANNED WITH AN ONGOING TRAINING REVIEW FOR CLEANING PROCESSES. BD WILL CONTINUE TO TREND COMPLAINTS FOR CONTAMINATION. H3 OTHER TEXT : SEE H.10.
IT WAS REPORTED THAT WHILE USING 10 BD BBL¿ TRYPTICASE¿ SOY AGAR WITH 5% SHEEP BLOOD (TSA II) // MACCONKEY II AGAR CONTAMINATION WAS OBSERVED BY THE LABORATORY PERSONNEL. THERE WAS NO INDICATION THAT RESULTS WERE REPORTED OUT AND THERE WAS NO REPORT OF PATIENT IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER "WE OBSERVED CONTAMINATION IN ABOUT 10 BIPLATES TSA II SB/MACC II WHEN THE PACKAGE WAS OPENED.".
THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED AS A DEFAULT A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED, BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED.
IT WAS REPORTED THAT WHILE USING 10 BD BBL¿ TRYPTICASE¿ SOY AGAR WITH 5% SHEEP BLOOD (TSA II) // MACCONKEY II AGAR CONTAMINATION WAS OBSERVED BY THE LABORATORY PERSONNEL. THERE WAS NO INDICATION THAT RESULTS WERE REPORTED OUT AND THERE WAS NO REPORT OF PATIENT IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER "WE OBSERVED CONTAMINATION IN ABOUT 10 BIPLATES TSA II SB/MACC II WHEN THE PACKAGE WAS OPENED."