COLLEAGUE CX VOLUMETRIC INFUSION PUMP
MDR report key: 1232852
·
Received November 18, 2008
Report
- Report Number
- 6000001-2007-05781
- Event Type
- Malfunction
- Date Received
- November 18, 2008
- Date of Event
- April 10, 2007
- Report Date
- April 10, 2007
- Manufacturer
- BAXTER HEALTHCARE PTE. LTD.
- Product Code
- FRN
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
EVALUATION SUMMARY: DURING PRODUCT EVALUATION, THE AIR IN LINE PRINTED CIRCUIT BOARD (AIL PCB) VALUES WERE CONFIRMED TO BE OUT OF SPECIFICATION. THE AIL PCB WAS RECALIBRATED.
Description of Event or Problem · 1
DURING PRODUCT EVALUATION THE PUMP'S AIR-IN-LINE WAS FOUND TO BE OUT OF SPECIFICATION. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION NECESSARY ACCORDING TO THE HOSPITAL REPRESENTATIVE. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | COLLEAGUE CX VOLUMETRIC INFUSION PUMP | INFUSION PUMP | FRN | BAXTER HEALTHCARE PTE. LTD. | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |