FDA Adverse Event Malfunction Summary report: N

BD VACUTAINER® K2 EDTA (K2E) 7.2MG BLOOD COLLECTION TUBES

MDR report key: 12327979 · Received August 16, 2021

Report

Report Number
1917413-2021-00738
Event Type
Malfunction
Date Received
August 16, 2021
Date of Event
July 22, 2021
Report Date
October 20, 2021
Manufacturer
BECTON, DICKINSON & CO. (BROKEN BOW)
Product Code
JKA
UDI-DI
50382903678612
PMA / PMN Number
BK050036
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

INVESTIGATION SUMMARY: BD HAD NOT RECEIVED SAMPLES OR PHOTOS FOR EVALUATION. THEREFORE, RETENTION SAMPLES OF THE INCIDENT LOT WERE SELECTED FROM BD INVENTORY FOR EVALUATION AND UPON COMPLETION, NO ISSUES RELATING TO MICROCLOTS WERE OBSERVED. 5 PRODUCTION LOT IN-HOUSE RETENTION SAMPLES FROM EACH LOT AFFECTED WERE VISUALLY INSPECTED WITH NO ISSUES BEING IDENTIFIED AND SUBMITTED TO THE CHEMISTRY LAB FOR TESTING. ALL RESULTS WERE WITHIN THE SPECIFICATION LIMITS OF 5.32MG ¿ 9.72MG FOR CATALOG 367861. NO DIFFICULTIES WERE ENCOUNTERED DURING BLOOD COLLECTION AND ALL TUBES APPEARED TO EXHIBIT PROPER FILL. BD WAS UNABLE TO CONFIRM THE CUSTOMER¿S INDICATED FAILURE (MICROCLOTS) BECAUSE THE DEFECT WAS NOT EVIDENT IN THE TESTING OF THE COMPLAINT LOT SAMPLES. LABORATORY ANALYSIS OF RETAIN AND CONTROL SAMPLES DEMONSTRATED CLINICALLY ACCEPTABLE PERFORMANCE FOR ALL VISUAL OBSERVATIONS EVALUATED. THE EDTA TUBES PERFORMED AS EXPECTED. BASED ON A REVIEW OF THE DEVICE HISTORY RECORD FOR THE INCIDENT LOT, ALL PRODUCT SPECIFICATIONS AND REQUIREMENTS FOR LOT RELEASE WERE MET. THERE WERE NO RELATED QUALITY ISSUES DURING MANUFACTURING OF THE PRODUCT. THIS COMPLAINT IS UNABLE TO BE CONFIRMED FOR THE INDICATED FAILURE MODE MICROCLOTS. BD WAS NOT ABLE TO IDENTIFY A ROOT CAUSE FOR THE INDICATED FAILURE MODE.

Description of Event or Problem · 0

IT WAS REPORTED WHEN USING THE BD VACUTAINER® K2 EDTA (K2E) 7.2MG BLOOD COLLECTION TUBES THERE WAS CLOTTING/MICRO CLOTS/FIBRIN CLOTS. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER. THE CUSTOMER STATED: "WE ARE EXPERIENCING CLOTTING WITH THE PRODUCT."

Additional Manufacturer Narrative · 1

THERE WERE MULTIPLE LOT NUMBERS REPORTED TO BE INVOLVED. THE INFORMATION FOR EACH LOT NUMBER IS AS FOLLOWS: MEDICAL DEVICE LOT #: 1040950, MEDICAL DEVICE EXPIRATION DATE: 2022-06-30, DEVICE MANUFACTURE DATE: 2021-02-09. MEDICAL DEVICE LOT #: 0316362. MEDICAL DEVICE EXPIRATION DATE: 2022-03-31. DEVICE MANUFACTURE DATE: 2020-11-11. A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 1

IT WAS REPORTED WHEN USING THE BD VACUTAINER® K2 EDTA (K2E) 7.2MG BLOOD COLLECTION TUBES THERE WAS CLOTTING/MICRO CLOTS/FIBRIN CLOTS. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER. THE CUSTOMER STATED: "WE ARE EXPERIENCING CLOTTING WITH THE PRODUCT."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1225821 BD VACUTAINER® K2 EDTA (K2E) 7.2MG BLOOD COLLECTION TUBES BLOOD SPECIMEN COLLECTION DEVICE JKA BECTON, DICKINSON & CO. (BROKEN BOW) 367861 SEE H.10 50382903678612

Patients

Seq Age Sex Outcome Treatment
1 Unknown