ENDOWRIST PROGRASP INSTRUMENT
Report
- Report Number
- 2955842-2008-01380
- Event Type
- Malfunction
- Date Received
- November 14, 2008
- Report Date
- November 14, 2008
- Manufacturer
- INTUITIVE SURGICAL, INC.
- Product Code
- GCJ
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- DC, US
- Reporter Occupation
- OTHER
Narratives
THE INSTRUMENT WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT, ENGINEERING CONFIRMED THE DISTAL END OF THE MAIN TUBE HAS A 5.750 INCH AREA, EXTENDED FROM THE INTERFACE BETWEEN THE PROXIMAL CLEVIS AND THE MAIN TUBE, WHERE MATERIAL HAS BEEN REMOVED FROM DIFFERENT SECTIONS OF THE TUBE. THE DAMAGED AREA IS PARALLEL TO THE AXIS AND HAS A ROUGH SURFACE FINISH. BASED ON THE LOCATION AND APPEARANCE, THE DAMGE WAS MOST LIKELY CAUSED BY A CANNULA ACCESSORY. DEEP SCRATCHES WERE ALSO OBSERVED CONCENTRATED WITHIN 2 INCHES FROM THE PROXIMAL CLEVIS WHICH MAY HAVE BEEN CAUSED BY INSTRUMENT COLLISIONS. ADDITIONAL INVESTIGATION IS UNDERWAY REGARDING THE CANNULA ACCESSORIES. A FOLLOW-UP MDR WILL BE SUBMITTED IF ADDITIONAL INFORMATION IS RECEIVED. NO OTHER DAMAGE FOUND. THE ENDOWRIST INSTRUMENTS INSTRUCTIONS FOR USE (IFU) SPECIFICALLY STATES: GENERAL PRECAUTIONS ANS WARNINGS: HANDLE INSTRUMENTS WITH CARE. AVOID MECHANICAL SHOCK OR STRESS THAT CAN CAUSE DAMAGE TO THE INSTRUMENTS. DO NOT USE AN INSTRUMENT TO CLEAN DEBRIS FROM ANOTHER INSTRUMENT INTRAOPERATIVELY. THIS MAY RESULT IN DAMGE TO THE INSTRUMENTS OR OTHER UNINTENDED CONSEQUENCES, SUCH AS DISCONNECTION OF THE INSTRUMENT TIP.
IT WAS REPORTED THAT THE ENDOWRIST PROGRASP INSTRUMENT IS SPLINTERING ON THE SHAFT. NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ENDOWRIST PROGRASP INSTRUMENT | ENDOSCOPIC INSTRUMENT | GCJ | INTUITIVE SURGICAL, INC. | 420093-07 | 0712071 345 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | AN ELECTROSURGICAL UNIT| DA VINCI S SURGICAL SYSTEM| ACCESSORIES |