FDA Adverse Event
Malfunction
Summary report: N
9800
MDR report key: 1232565
·
Received November 13, 2008
Report
- Report Number
- 1720753-2008-27993
- Event Type
- Malfunction
- Date Received
- November 13, 2008
- Date of Event
- October 23, 2008
- Report Date
- November 12, 2008
- Manufacturer
- GE OEC MEDICAL SYSTEMS INC.
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- VA, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
THIS MDR IS RELATED TO GE HEALTHCARE COMPLAINT NUMBER. THE GE SERVICE REP FOUND THE SLOW SYSTEM RESPONSE WAS DUE TO THE SYSTEM HARD DRIVE. HE REMOVED AND REPLACED THE SYSTEM HARD DRIVE, AND LOADED JV 5.5 REL 29 SOFTWARE. THE SYSTEM OPERATES AS INTENDED.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT THE SYSTEM DISPLAYED A SLOW RESPONSE. NO PATIENT INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | 9800 | FLUOROSCOPIC X-RAY | JAA | GE OEC MEDICAL SYSTEMS INC. | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |