FDA Adverse Event Malfunction Summary report: N

ID NOW COVID-19

MDR report key: 12321581 · Received August 13, 2021

Report

Report Number
1221359-2021-02261
Event Type
Malfunction
Date Received
August 13, 2021
Report Date
August 13, 2021
Manufacturer
ABBOTT DIAGNOSTICS SCABOROUGH
Product Code
QJR
UDI-DI
10811877011320
PMA / PMN Number
EUA2000074
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
ID
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

INVESTIGATION: TESTING WAS PERFORMED AT ABBOTT DIAGNOSTICS (B)(4), INC. ON RETAINED KIT LOT M140253 WITH INTERNAL POSITIVE QUALITY CONTROL SAMPLES AND NEGATIVE QUALITY CONTROL SWABS. ALL TEST RESULTS WERE VALID AND PERFORMED AS EXPECTED. ADDITIONALLY, THE MANUFACTURING RECORDS AND QUALITY CONTROL RELEASE TESTING WAS REVIEWED FOR TEST KIT TEST BASE PART NUMBER 190-430 / LOT M140253. THE LOT MET THE REQUIRED RELEASE SPECIFICATIONS. A REVIEW OF THE COMPLAINTS REPORTED AS FALSE POSITIVE PATIENT RESULTS (CONFIRMED AND UNCONFIRMED, CONFLICTING RESULTS) RELATED TO KIT LOT (B)(4) ABBOTT DIAGNOSTICS (B)(4) WAS UNABLE TO DETERMINE THE EXACT ROOT CAUSE OF THE REPORTED ISSUE HOWEVER A POSSIBLE ASSIGNABLE ROOT CAUSE IS SAMPLE INTERFERENCE OR CROSS CONTAMINATION. THIS EVENT WAS ORIGINALLY REPORTED IN FOLLOW MDR # 1 1221359-2021-01872. REFERENCE MFR NUMBERS: 1221359-2021-01877, 1221359-2021-02260, 1221359-2021-02262.

Description of Event or Problem · 0

THE CUSTOMER REPORTED A DISCREPANT POSITIVE TEST RESULT FOR ID NOW COVID-19 THAT TESTED NEGATIVE ON PCR TEST. THE CUSTOMER REPORTED THAT THE PATIENT WAS SYMPTOMATIC PRESENTING TO THE EMERGENCY ROOM FEVER AND HEADACHE. THE PATIENT RECEIVED POSITIVE RESULTS WITH ID NOW. PCR CONFIRMATION TESTING GENERATED NEGATIVE RESULTS. THE PULMONOLOGIST HAD TAKEN THE PATIENT FOR X-RAY TEST, BUT THE CLINICAL PATHOLOGIST DID NOT KNOW THE RESULT WHETHER THIS PATIENT HAD PNEUMONIA OR NOT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1214784 ID NOW COVID-19 MOLECULAR IVD FOR ID NOW COVID-19 QJR ABBOTT DIAGNOSTICS SCABOROUGH M140253 10811877011320

Patients

Seq Age Sex Outcome Treatment
1