FDA Adverse Event Malfunction Summary report: N

CENTRA BED

MDR report key: 1232149 · Received November 4, 2008

Report

Report Number
1824206-2008-04038
Event Type
Malfunction
Date Received
November 4, 2008
Date of Event
October 7, 2008
Report Date
October 7, 2008
Manufacturer
HILL-ROM RITTER
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE HILL-ROM TECHNICIAN FOUND THE METAL TAB ON THE REVERSE TRENDELENBURG SWITCH WAS BENT. THE TECHNICIAN STRAIGHTENED THE TAB TO REPAIR THE BED.

Description of Event or Problem · 1

INFORMATION RECEIVED INDICATES THE BED WILL NOT GO INTO THE REVERSE TRENDELENBURG POSITION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CENTRA BED AC POWERED HOSPITAL BED FNL HILL-ROM RITTER 850

Patients

Seq Age Sex Outcome Treatment
1