FDA Adverse Event
Malfunction
Summary report: N
4 MOTOR ADVANCE BED
MDR report key: 1232123
·
Received November 4, 2008
Report
- Report Number
- 1824206-2008-04044
- Event Type
- Malfunction
- Date Received
- November 4, 2008
- Date of Event
- October 8, 2008
- Report Date
- October 8, 2008
- Manufacturer
- HILL-ROM RITTER
- Product Code
- FNL
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MO, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE FACILITIES ENGINEER FOUND THE P73 CONNECTOR ON THE KNEE LIMIT CABLE RIBBON CARRIER WAS WIRED BACKWARDS. THE ENGINEER CORRECTED THE WIRING TO REPAIR THE BED.
Description of Event or Problem · 1
INFORMATION RECEIVED INDICATES THE KNEE FUNCTION MOVED UP UNINTENTIONALLY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | 4 MOTOR ADVANCE BED | AC POWERED HOSPITAL BED | FNL | HILL-ROM RITTER | 1104 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |