FDA Adverse Event Malfunction Summary report: N

4 MOTOR ADVANCE BED

MDR report key: 1232123 · Received November 4, 2008

Report

Report Number
1824206-2008-04044
Event Type
Malfunction
Date Received
November 4, 2008
Date of Event
October 8, 2008
Report Date
October 8, 2008
Manufacturer
HILL-ROM RITTER
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MO, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE FACILITIES ENGINEER FOUND THE P73 CONNECTOR ON THE KNEE LIMIT CABLE RIBBON CARRIER WAS WIRED BACKWARDS. THE ENGINEER CORRECTED THE WIRING TO REPAIR THE BED.

Description of Event or Problem · 1

INFORMATION RECEIVED INDICATES THE KNEE FUNCTION MOVED UP UNINTENTIONALLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 4 MOTOR ADVANCE BED AC POWERED HOSPITAL BED FNL HILL-ROM RITTER 1104

Patients

Seq Age Sex Outcome Treatment
1