FDA Adverse Event
Malfunction
Summary report: N
TOTALCARE
MDR report key: 1232046
·
Received November 7, 2008
Report
- Report Number
- 1824206-2008-04102
- Event Type
- Malfunction
- Date Received
- November 7, 2008
- Date of Event
- October 10, 2008
- Report Date
- October 10, 2008
- Manufacturer
- HILL-ROM RITTER
- Product Code
- FNL
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MS, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE HILL-ROM TECHNICIAN INVESTIGATED AND DISCOVERED WHEN THE HI/LOW WAS LOWERED TO THE LOW LIMIT, THE SCALE WAS WEIGHING OFF BY 30 LBS. THE TECHNICIAN FOUND THE LOAD BEAM WIRES WERE BEING PULLED TIGHT WHEN THE BED WAS LOWERED, WHICH CAUSED THE INACCURATE WEIGHT. THE TECHNICIAN PROPERLY ROUTED THE LOAD BEAM WIRES TO REPAIR THE BED.
Description of Event or Problem · 1
INFORMATION RECEIVED INDICATES THE BED IS ALARMING.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | TOTALCARE | AC POWERED HOSPITAL BED | FNL | HILL-ROM RITTER | 1900 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |