PUMP MMT-1712KL 640G V4.10 BK SF MG
Report
- Report Number
- 2032227-2021-180072
- Event Type
- Malfunction
- Date Received
- August 12, 2021
- Date of Event
- August 3, 2021
- Report Date
- May 2, 2022
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- UDI-DI
- 000000763000367046
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IT
- Reporter Occupation
- OTHER
- Health Professional
- N
Narratives
RETAINER RING = BLACK CUSTOMER RETURNED INSULIN PUMP FOR AN ALLEGED CRITICAL PUMP ERROR (OPEN BOOK IMAGE) ALARM FOUND ON (B)(6) 2021. INSULIN PUMP WAS RECEIVED WITH A CRITICAL PUMP ERROR (OPEN BOOK IMAGE) ALARM. UNABLE TO PERFORM THE DISPLACEMENT TEST, REWIND TEST, PRIME/SEATING TEST, BASIC OCCLUSION TEST, FORCE SENSOR TEST, OCCLUSION TEST, SLEEP CURRENT MEASUREMENT TEST, ACTIVE CURRENT MEASUREMENT TEST AND SELF TEST DUE TO CRITICAL PUMP ERROR (OPEN BOOK IMAGE) ALARM. INSULIN PUMP WAS CUT OPEN TO PERFORM VISUAL INSPECTION AND FOUND NO EVIDENCE OF PHYSICAL OR MOISTURE DAMAGE ON THE ELECTRICAL BOARD, FORCE SENSOR, MOTOR AND VIBRATOR ASSEMBLY NOTED. SUCCESSFULLY UTILIZED CREST AND THUS TO DOWNLOAD HISTORY FILES, TRACES AND COMLINK3 FILES. CRITICAL PUMP ERROR (OPEN BOOK IMAGE) ALARM TRIGGERED BY THREE CONSECUTIVE PUMP ERROR 53 ALARMS ON (B)(6) 2021 06:24:20.000, (B)(6) 2021 06:27:12.000, (B)(6) 2021 06:28:02.000, (B)(6) 2021 06:29:03.000, (B)(6) 2021 06:33:34.000,(B)(6) 2021 06:36:04.000 AND (B)(6) 2021 06:40:15.000. PUMP ERROR 53 ALARM DUE TO SOFTWARE ERROR (LINENUMBER = (B)(4) ; FILENUMBER = (B)(4)). FORCE SENSOR ZERO OFFSET WITHIN SPECIFICATION (23.5 MV). TEST P-CAP AND RESERVOIR LOCKED PROPERLY INTO RESERVOIR COMPARTMENT DURING TESTING. THE FOLLOWING WERE NOTED DURING VISUAL INSPECTION: A PILLOWING KEYPAD OVERLAY, A CRACKED BATTERY TUBE THREADS, A SCRATCHED CASE, A CRACKED CASE-CORNER OF BELT CLIP RAILS NEAR THE BATTERY TUBE COMPARTMENT, A SERIAL NUMBER LABEL FADING AND A BROKEN BELT CLIP RAILS. CRITICAL PUMP ERROR (OPEN BOOK IMAGE) ALARM CONFIRMED DUE TO THREE CONSECUTIVE PUMP ERROR 53 ALARMS. PUMP ERROR 53 ALARMS DUE TO SOFTWARE ERROR. MEDTRONIC, INC. (MEDTRONIC) IS SUBMITTING THIS REPORT TO COMPLY WITH 21 C.F.R. PART 803, THE MEDICAL DEVICE REPORTING REGULATION. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION IN THE TIME ALLOTTED AND HAS PROVIDED AS MUCH INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY "DEFECTS" OR HAS "MALFUNCTIONED". THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA, TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REPORTING PURSUANT TO PART 803. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE IT BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT.
(B)(4). CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT INSULIN PUMP HAD CRITICAL PUMP ERROR ALARM. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE CUSTOMER WAS ASSISTED WITH TROUBLESHOOTING. THE DEVICE WILL BE RETURNED FOR ANALYSIS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1214233 | PUMP MMT-1712KL 640G V4.10 BK SF MG | PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-1712KL | HG4QSCU | 000000763000367046 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |