FDA Adverse Event
Malfunction
Summary report: N
COLLEAGUE 3 VOLUMETRIC INFUSION PUMP
MDR report key: 1231466
·
Received November 17, 2008
Report
- Report Number
- 6000001-2007-05569
- Event Type
- Malfunction
- Date Received
- November 17, 2008
- Date of Event
- March 23, 2007
- Report Date
- March 23, 2007
- Manufacturer
- BAXTER HEALTHCARE PTE. LTD.
- Product Code
- FRN
- Removal / Correction Number
- 2005-047-MD
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- LA, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
EVALUATION SUMMARY: THE CONDITION OF A DEFECTIVE USER INTERFACE MODULE PRINTED CIRCUIT BOARD (UIM PCB) WAS CONFIRMED. FAILURE CODE 703 IN THE EVENT HISTORY CONFIRMS THE DEFECTIVE UIM PCB. THIS FAILURE CODE IS MANIFESTED AS A RESULT OF THE UIM PCB BEING DEFECTIVE. THE UIM PCB WAS REPLACED. REVIEW OF THE COMPLAINT HISTORY REVEALS SIMILAR REPORTS HAVE BEEN RECEIVED FOR THIS PRODUCT FOR THE REPORTED ISSUE. THIS ISSUE IS BEING INVESTIGATED.
Description of Event or Problem · 1
DURING PRODUCT EVALUATION THE PUMP'S USER INTERFACE MODULE PRINTED CIRCUIT BOARD (UIM PCB) WAS FOUND TO BE DEFECTIVE. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION NECESSARY ACCORDING TO THE HOSPITAL REPRESENTATIVE. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | COLLEAGUE 3 VOLUMETRIC INFUSION PUMP | INFUSION PUMP | FRN | BAXTER HEALTHCARE PTE. LTD. | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |