FLO-GARD 6201 VOLUMETRIC INFUSION PUMP
Report
- Report Number
- 6000001-2007-88504
- Event Type
- Malfunction
- Date Received
- November 17, 2008
- Date of Event
- June 1, 2007
- Report Date
- June 1, 2007
- Manufacturer
- BAXTER HEALTHCARE (SG)
- Product Code
- FRN
- PMA / PMN Number
- K915522
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CO, US
- Reporter Occupation
- OTHER
Narratives
THE FACILITY REPRESENTATIVE REPORTED ¿CONSTANT ALARM AND THE PUMP WILL NOT RUN¿. INFORMATION WAS NOT PROVIDED REGARDING WHETHER OR NOT THE PROBLEM OCCURRED DURING A PATIENT INFUSION. EVALUATION SUMMARY: A BAXTER SERVICE TECHNICIAN EVALUATED THE DEVICE AND FOUND A CRACKED DOOR ASSEMBLY. THE REPORTED CONDITION OF ¿CONSTANT ALARM AND WILL NOT RUN¿ WAS CONFIRMED, CAUSED BY A CRACKED DOOR ASSEMBLY. REVIEW OF THE COMPLAINT HISTORY REVEALS SIMILAR REPORTS HAVE BEEN RECEIVED FOR THIS PRODUCT FOR THE REPORTED ISSUE. THIS ISSUE IS BEING INVESTIGATED UNDER CAPA.
THE DEVICE WAS RETURNED FOR SERVICE. DURING SERVICE BY BAXTER, A CRACKED DOOR ASSEMBLY WAS FOUND. ACCORDING TO THE FACILITY REPRESENTATIVE, NO PATIENT INJURY OR MEDICAL INTERVENTION HAD BEEN REPORTED RELATED TO THE DEVICE. NO ADDITIONAL INFORMATION OR CONTACT WAS AVAILABLE.