FLO-GARD 6201 VOLUMETRIC INFUSION PUMP
MDR report key: 1231283
·
Received November 17, 2008
Report
- Report Number
- 6000001-2007-88293
- Event Type
- Malfunction
- Date Received
- November 17, 2008
- Date of Event
- May 1, 2007
- Report Date
- May 11, 2007
- Manufacturer
- BAXTER HEALTHCARE (SG)
- Product Code
- FRN
- PMA / PMN Number
- K915522
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
Additional Manufacturer Narrative · 1
THE FACILITY REPRESENTATIVE REPORTED A CONSTANT OCCLUSION ALARM FOUND DURING BIO-MED TESTING. EVALUATION SUMMARY: THE PUMP WAS EVALUATED BY A BAXTER SERVICE TECHNICIAN. INSPECTION OF THE DEVICE FOUND A DOWNSTREAM OCCLUSION ALARM CAUSED BY A BROKEN DOOR ASSEMBLY. REVIEW OF THE COMPLAINT HISTORY REVEALS SIMILAR REPORTS HAVE BEEN RECEIVED FOR THIS PRODUCT FOR THE REPORTED ISSUE. THIS ISSUE IS BEING INVESTIGATED.
Description of Event or Problem · 1
THE DEVICE WAS RETURNED FOR SERVICE. DURING SERVICE BY BAXTER, A BROKEN DOOR ASSEMBLY WAS FOUND. ACCORDING TO THE HOSPITAL REPRESENTATIVE, NO PATIENT INJURY OR MEDICAL INTERVENTION HAD BEEN REPORTED RELATED TO THE DEVICE. NO ADDITIONAL INFORMATION IS AVAILABLE.