FDA Adverse Event Malfunction Summary report: N

FLO-GARD 6201 VOLUMETRIC INFUSION PUMP

MDR report key: 1231283 · Received November 17, 2008

Report

Report Number
6000001-2007-88293
Event Type
Malfunction
Date Received
November 17, 2008
Date of Event
May 1, 2007
Report Date
May 11, 2007
Manufacturer
BAXTER HEALTHCARE (SG)
Product Code
FRN
PMA / PMN Number
K915522
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

THE FACILITY REPRESENTATIVE REPORTED A CONSTANT OCCLUSION ALARM FOUND DURING BIO-MED TESTING. EVALUATION SUMMARY: THE PUMP WAS EVALUATED BY A BAXTER SERVICE TECHNICIAN. INSPECTION OF THE DEVICE FOUND A DOWNSTREAM OCCLUSION ALARM CAUSED BY A BROKEN DOOR ASSEMBLY. REVIEW OF THE COMPLAINT HISTORY REVEALS SIMILAR REPORTS HAVE BEEN RECEIVED FOR THIS PRODUCT FOR THE REPORTED ISSUE. THIS ISSUE IS BEING INVESTIGATED.

Description of Event or Problem · 1

THE DEVICE WAS RETURNED FOR SERVICE. DURING SERVICE BY BAXTER, A BROKEN DOOR ASSEMBLY WAS FOUND. ACCORDING TO THE HOSPITAL REPRESENTATIVE, NO PATIENT INJURY OR MEDICAL INTERVENTION HAD BEEN REPORTED RELATED TO THE DEVICE. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 FLO-GARD 6201 VOLUMETRIC INFUSION PUMP INFUSION PUMP FRN BAXTER HEALTHCARE (SG) NA NA

Patients

Seq Age Sex Outcome Treatment
1