FDA Adverse Event
Malfunction
Summary report: N
FG 6301 DUAL CHNL VOL INF PUMP2M8064 SERVICE
MDR report key: 1231263
·
Received November 17, 2008
Report
- Report Number
- 6000001-2007-88297
- Event Type
- Malfunction
- Date Received
- November 17, 2008
- Date of Event
- May 1, 2007
- Report Date
- May 10, 2007
- Manufacturer
- BAXTER HEALTHCARE (SG)
- Product Code
- FRN
- PMA / PMN Number
- K915523
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
EVALUATION SUMMARY: A BAXTER SERVICE TECHNICIAN EVALUATED THE DEVICE AND FOUND CRACKED DOORS #1 AND #2. THE REPORTED CONDITION OF THE BROKEN DOOR WAS CONFIRMED. REVIEW OF THE COMPLAINT HISTORY REVEALS SIMILAR REPORTS HAVE BEEN RECEIVED FOR THIS PRODUCT FOR THE REPORTED ISSUE. THIS ISSUE IS BEING INVESTIGATED.
Description of Event or Problem · 1
THE FACILITY REPRESENTATIVE REPORTED A BROKEN DOOR DURING BIO-MED TESTING. ALTHOUGH BAXTER ATTEMPTED TO OBTAIN INFORMATION, DETAILS WERE NOT AVAILABLE REGARDING ADDITIONAL CONTACT INFORMATION. ACCORDING TO THE FACILITY REPRESENTATIVE, NO PATIENT INJURY OR MEDICAL INTERVENTION HAD BEEN REPORTED RELATED TO THE DEVICE. NO ADDITIONAL INFORMATION WAS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | FG 6301 DUAL CHNL VOL INF PUMP2M8064 SERVICE | INFUSION PUMP | FRN | BAXTER HEALTHCARE (SG) | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |