FDA Adverse Event Injury Summary report: N

COMANECI

MDR report key: 12312394 · Received August 12, 2021

Report

Report Number
3015531723-2021-00005
Event Type
Injury
Date Received
August 12, 2021
Date of Event
July 30, 2021
Report Date
August 12, 2021
Manufacturer
RAPID MEDICAL LTD.
Product Code
PUU
UDI-DI
07290015107150
Adverse Event
Yes
Report Source
Distributor report
Reporter Location
FL, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE HAS NOT BEEN RETURNED FOR EVALUATION. THEREFORE, THE ROOT CAUSE OF THE EVENT CANNOT BE DETERMINED. COIL PROTRUSION IS A KNOWN POTENTIAL COMPLICATION ASSOCIATED WITH THE USE OF COIL EMBOLIZATION ASSIST DEVICES, INCLUDING COMANECI. HOWEVER, ACCORDING TO PHYSICIAN REPORT, IT IS POSSIBLE THAT THE PROTRUDING OF THE SUBJECT DEVICE INTO THE ANEURYSM, AS IT WAS BEING WITHDRAWN, MAY HAVE BEEN ASSOCIATED TO THE PATIENT'S CHALLENGING ACUTE BASILAR-PCA ANGLE.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING TREATMENT OF A WIDE-NECKED BASILAR TIP ANEURYSM, THE COMANECI 17 WAS PROTRUDED INTO THE ANEURYSM COIL MASS AS IT WAS BEING WITHDRAWN. AFTER MULTIPLE MANIPULATIONS THE PHYSICIAN WITHDREW THE DEVICE AND COMPLETED THE ANEURYSM COILING PROCEDURE, HOWEVER ONE COIL REMAINED HERNIATED INTO THE PARENT ARTERY. THERE WAS NO REPORTED PATIENT INJURY OR CLINICAL CONSEQUENCE. THE PATIENT IS REPORTED TO BE DOING WELL.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1212489 COMANECI COMANECI 17 PUU RAPID MEDICAL LTD. ANPP7199 210304CS01 07290015107150

Patients

Seq Age Sex Outcome Treatment
1 Disability