FDA Adverse Event Malfunction Summary report: N

ETS FLEX ARTICING LNR CUTR 45MM

MDR report key: 1231203 · Received November 10, 2008

Report

Report Number
3005075853-2008-02905
Event Type
Malfunction
Date Received
November 10, 2008
Report Date
October 10, 2008
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
GDW
PMA / PMN Number
K020779
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

DAMAGED ARTICULATION GEAR TEETH. EVALUATION SUMMARY: THE ANALYSIS SHOWED THAT THE DEVICE WAS RECEIVED WITH THE ARTICULATION MECHANISM DAMAGED. THE DEVICE WAS RECEIVED WITH TWO RELOADS PRESENT; THE RELOADS WERE RECEIVED FULLY FIRED AND WITH NO DAMAGE TO THE SPRING RELOAD LOCKOUT. THE RETURNED DEVICE WAS TESTED FOR FUNCTIONALITY WITH A TEST RELOAD ON THE THREE ARTICULATED POSITIONS; WHEN FIRED TO THE LEFT AND CENTER, THE STAPLE FORMATION WAS CONFORMING. HOWEVER, WHEN FIRED IN THE RIGHT POSITION, THE STAPLES WERE MALFORMED DUE TO THE DAMAGED ARTICULATION MECHANISM. THE DEVICE WAS DISASSEMBLED TO VERIFY THE CONDITION OF THE INTERNAL COMPONENTS AND THE ARTICULATION GEAR TEETH WERE FOUND BROKEN. IT SHOULD BE NOTED THAT A 100% INSPECTIONS TAKES PLACE DURING MANUFACTURING TO ENSURE THE DEVICE MEETS THE REQUIRED SPECIFICATIONS; IN ADDITION, A SAMPLE OF THE BATCH IS INSPECTED AT FGQA. WE HAVE DOCUMENTED THE CIRCUMSTANCES AS THEY WERE REPORTED TO US. IN ADDITION, THE APPROPRIATE ENGINEERING PERSONNEL HAVE BEEN NOTIFIED OF THIS INCIDENT. AN INVESTIGATION HAS BEEN INITIATED TO DETERMINE THE CAUSE OF THIS ISSUE. A BATCH RECORD REVIEW WAS PERFORMED AND NO ANOMALIES WERE FOUND DURING THE MANUFACTURING PROCESS.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A COLECTOMY, THE DEVICE LOCKED OUT. IT WOULD NOT FIRE AT ALL. ANOTHER DEVICE WAS USED TO COMPLETE THE CASE. THERE WAS NO ADVERSE CONSEQUENCE TO THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ETS FLEX ARTICING LNR CUTR 45MM GDW ETHICON ENDO-SURGERY, LLC. NA UNK