FG 6301 DUAL CHNL VOL INF PUMP2M8064 SERVICE
Report
- Report Number
- 6000001-2007-88349
- Event Type
- Malfunction
- Date Received
- November 17, 2008
- Date of Event
- April 1, 2007
- Report Date
- April 27, 2007
- Manufacturer
- BAXTER HEALTHCARE (SG)
- Product Code
- FRN
- PMA / PMN Number
- K915523
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
EVALUATION SUMMARY: THE PUMP WAS EVALUATED BY A BAXTER SERVICE TECHNICIAN. THE REPORTED CONDITION OF ¿UNKNOWN¿ WAS CAUSED BY A BROKEN DOOR. REVIEW OF THE COMPLAINT HISTORY REVEALS SIMILAR REPORTS HAVE BEEN RECEIVED FOR THIS PRODUCT FOR THE REPORTED ISSUE. THIS ISSUE IS BEING INVESTIGATED.
THE DEVICE WAS RETURNED FOR SERVICE. DURING SERVICE BY BAXTER, CRACKED DOORS #1 AND #2 WERE FOUND. ACCORDING TO THE FACILITY REPRESENTATIVE, NO PATIENT INJURY OR MEDICAL INTERVENTION HAD BEEN REPORTED RELATED TO THE DEVICE. NO ADDITIONAL INFORMATION OR CONTACT WAS AVAILABLE.
THE FACILITY REPRESENTATIVE REPORTED ¿UNKNOWN¿. INFORMATION WAS NOT PROVIDED REGARDING WHETHER OR NOT THE PROBLEM OCCURRED DURING A PATIENT INFUSION. ALTHOUGH BAXTER ATTEMPTED TO OBTAIN INFORMATION, DETAILS WERE NOT AVAILABLE REGARDING ADDITIONAL CONTACT INFORMATION. ACCORDING TO THE FACILITY REPRESENTATIVE, NO PATIENT INJURY OR MEDICAL INTERVENTION HAD BEEN REPORTED RELATED TO THE DEVICE. NO ADDITIONAL INFORMATION WAS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | FG 6301 DUAL CHNL VOL INF PUMP2M8064 SERVICE | INFUSION PUMP | FRN | BAXTER HEALTHCARE (SG) | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |