STERRAD NX STERILIZER
Report
- Report Number
- 2084725-2008-00705
- Event Type
- Malfunction
- Date Received
- October 30, 2008
- Date of Event
- October 9, 2008
- Report Date
- October 9, 2008
- Manufacturer
- ADVANCED STERILIZATION PRODUCTS
- Product Code
- MLR
- PMA / PMN Number
- K042116
- Removal / Correction Number
- Z-0844-2008
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IN, US
- Reporter Occupation
- OTHER
Narratives
DAMAGED DEVICE, CONCLUSION: THE REPORTED ISSUE HAS BEEN DOCUMENTED INTO ASP'S COMPLAINT SYSTEM FOR TRACKING AND TRENDING PURPOSES. AS MANUFACTURERS INTRODUCE NEW TECHNOLOGIES AND MAKE MATERIALS, MFG AND REPAIR PROCESS CHANGED TO THEIR DEVICES, ASP DOES NOT HAVE THE MEANS TO PROVIDED UP-TO-DATE INFO RELATED TO SPECIFIC BRANDS AND MODELS OF MEDICAL DEVICES THAT MAY BE PROCESSED IN THE STERRAD SYSTEMS. ASP MAY WORK WITH SPECIFIC DEVICE MANUFACTURERS TO TEST FOR COMPATIBILITY IN THE STERRAD SYSTEMS, HOWEVER, THE COMPATIBILITY DATA AND THE DECISION AS TO WHETHER THE DEVICE IS COMPATIBLE IN THE STERRAD SYSTEMS IS ULTIMATELY THE DECISION OF THE DEVICE MFR. IN 2007, A CAPA WAS INITIATED AND CUSTOMER LETTER WAS SENT TO ALL STERRAD SYSTEMS USERS TO DIRECTLY CONTACT THE DEVICE MFR REGARDING MATERIAL COMPATIBILITY AND FUNCTIONAL PERFORMANCE QUESTIONS OF THE DEVICE WITH THE STERRAD SYSTEMS. ASP WILL CONTINUE TO TRACK AND TREND.
AN ASP ACCOUNT EXECUTIVE REPORTED THAT HE HAD A CUSTOMER WHO PROCESSED A RIGID ACMI SCOPE ON THE ADVANCED CYCLE INSTEAD OF THE STANDARD AND THE BLACK PART OF THE EYEPIECE BUBBLED. THE CUSTOMER INVESTIGATED THE DAMAGED SCOPE AND IT WAS DETERMINED THAT THE MR 6 SCOPE HAS A DUAL LUMEN. THERE WERE NO REPORTS OF ANY INJURY TO A PT OR A HEALTHCARE WORKER.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | STERRAD NX STERILIZER | STERRAD EQUIPMENT (MLR) | MLR | ADVANCED STERILIZATION PRODUCTS | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | ACMI SHORT RIGID UTEROSCOPE |