FDA Adverse Event Malfunction Summary report: N

STERRAD 200 STERILIZATION SINGLE-DOOR UNIT WITH STAINLESS STEEL EXTERIOR PANELS

MDR report key: 1231051 · Received October 30, 2008

Report

Report Number
2084725-2008-00714
Event Type
Malfunction
Date Received
October 30, 2008
Date of Event
October 7, 2008
Report Date
October 7, 2008
Manufacturer
YAMATO SCIENTIFIC CO.
Product Code
MLR
PMA / PMN Number
K030429
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AZ, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

OIL MIST - CAPITAL EQUIPMENT, EVALUATED AT CUSTOMER SITE. THE FSE SAW AND SMELLED OIL MIST. THE FSE REPLACED THE OIL MIST FILTER ASSEMBLY. THE FSE TESTED FOR OIL MIST OR SMELL. THERE WAS NONE FOUND. THE FSE RAN EMPTY LOAD, ALL PASSED.

Description of Event or Problem · 1

THE CUSTOMER REQUESTED THE OIL FILTER TO BE CHANGED. HE SAID IT SMELLED, BUT THERE WERE NO REPORTS OF WATERY EYES OR OTHER RELATED SYMPTOMS. THE ASP FIELD SERVICE ENGINEER (FSE) WENT TO THE FACILITY TO ASSESS THE UNIT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 STERRAD 200 STERILIZATION SINGLE-DOOR UNIT WITH STAINLESS STEEL EXTERIOR PANELS STERILIZER, MEDICAL DEVICE MLR YAMATO SCIENTIFIC CO. NA NA

Patients

Seq Age Sex Outcome Treatment
1 NA