FDA Adverse Event
Malfunction
Summary report: N
STERRAD 200 STERILIZATION SINGLE-DOOR UNIT WITH STAINLESS STEEL EXTERIOR PANELS
MDR report key: 1231051
·
Received October 30, 2008
Report
- Report Number
- 2084725-2008-00714
- Event Type
- Malfunction
- Date Received
- October 30, 2008
- Date of Event
- October 7, 2008
- Report Date
- October 7, 2008
- Manufacturer
- YAMATO SCIENTIFIC CO.
- Product Code
- MLR
- PMA / PMN Number
- K030429
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AZ, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
OIL MIST - CAPITAL EQUIPMENT, EVALUATED AT CUSTOMER SITE. THE FSE SAW AND SMELLED OIL MIST. THE FSE REPLACED THE OIL MIST FILTER ASSEMBLY. THE FSE TESTED FOR OIL MIST OR SMELL. THERE WAS NONE FOUND. THE FSE RAN EMPTY LOAD, ALL PASSED.
Description of Event or Problem · 1
THE CUSTOMER REQUESTED THE OIL FILTER TO BE CHANGED. HE SAID IT SMELLED, BUT THERE WERE NO REPORTS OF WATERY EYES OR OTHER RELATED SYMPTOMS. THE ASP FIELD SERVICE ENGINEER (FSE) WENT TO THE FACILITY TO ASSESS THE UNIT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | STERRAD 200 STERILIZATION SINGLE-DOOR UNIT WITH STAINLESS STEEL EXTERIOR PANELS | STERILIZER, MEDICAL DEVICE | MLR | YAMATO SCIENTIFIC CO. | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA |