FDA Adverse Event Malfunction Summary report: N

ANALYTICAL P MODULE

MDR report key: 1230992 · Received October 28, 2008

Report

Report Number
1823260-2008-07977
Event Type
Malfunction
Date Received
October 28, 2008
Date of Event
October 6, 2008
Report Date
October 28, 2008
Manufacturer
ROCHE DIAGNOSTICS
Product Code
JJE
PMA / PMN Number
K953239
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
BE
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Description of Event or Problem · 1

FOR BILIRUBIN, BICARBONATE AND ALBUMIN, FALSE RESULTS HAVE BEEN GENERATED FOR CONTROLS AND PT SAMPLES. AFTER RECALIBRATION, THE PHENOMENON DISAPPEARED. NO SAMPLE RESULTS HAVE BEEN REPORTED. THE INVESTIGATIONAL UNIT DETERMINED THE ROOT CAUSE IS A TIMING CONFLICT WHEN WRITING REAGENT AND CALIBRATION PARAMETER INFO TO THE HARD DISK. CUSTOMER WILL BE INFORMED BY FIELD SAFETY NOTICE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ANALYTICAL P MODULE CLINICAL CHEMISTRY ANALYZER-JJE JJE ROCHE DIAGNOSTICS

Patients

Seq Age Sex Outcome Treatment
1 UNK