FDA Adverse Event
Malfunction
Summary report: N
ANALYTICAL P MODULE
MDR report key: 1230992
·
Received October 28, 2008
Report
- Report Number
- 1823260-2008-07977
- Event Type
- Malfunction
- Date Received
- October 28, 2008
- Date of Event
- October 6, 2008
- Report Date
- October 28, 2008
- Manufacturer
- ROCHE DIAGNOSTICS
- Product Code
- JJE
- PMA / PMN Number
- K953239
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- BE
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
Description of Event or Problem · 1
FOR BILIRUBIN, BICARBONATE AND ALBUMIN, FALSE RESULTS HAVE BEEN GENERATED FOR CONTROLS AND PT SAMPLES. AFTER RECALIBRATION, THE PHENOMENON DISAPPEARED. NO SAMPLE RESULTS HAVE BEEN REPORTED. THE INVESTIGATIONAL UNIT DETERMINED THE ROOT CAUSE IS A TIMING CONFLICT WHEN WRITING REAGENT AND CALIBRATION PARAMETER INFO TO THE HARD DISK. CUSTOMER WILL BE INFORMED BY FIELD SAFETY NOTICE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ANALYTICAL P MODULE | CLINICAL CHEMISTRY ANALYZER-JJE | JJE | ROCHE DIAGNOSTICS |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |