MENTOR MEMORYGEL BREAST IMPLANT
Report
- Report Number
- 1645337-2021-08967
- Event Type
- Injury
- Date Received
- August 11, 2021
- Date of Event
- July 20, 2021
- Report Date
- July 20, 2021
- Manufacturer
- MENTOR TEXAS
- Product Code
- FTR
- UDI-DI
- 00081317000310
- PMA / PMN Number
- P030053
- Removal / Correction Number
- N/A
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CT, US
- Reporter Occupation
- OTHER
Narratives
ON AUGUST 24, 2021, THE MENTOR FAILURE ANALYSIS LAB RECEIVED THE DEVICE FOR EVALUATION. ON AUGUST 31, 2021, MENTOR COMPLETED EVALUATION OF THE DEVICE. DEVICE EVALUATION SUMMARY: THE PRODUCT WAS RETURNED TO MENTOR FOR EVALUATION. MENTOR CONDUCTED A VISUAL INSPECTION OF THE RETURNED DEVICE. VISUAL ANALYSIS OF THE RETURNED SAMPLE REVEALED THAT THE SM MPP GEL 375CC BREAST IMPLANT HAD A CREASE/FOLD ON THE POSTERIOR VIEW. ADDITIONALLY, A TEAR WAS NOTED WITHIN THE CREASE, MEASURING APPROXIMATELY 7.0 CM THE EVALUATION DETERMINED THAT THE CAUSE OF THE RUPTURE IS CONSISTENT WITH NORMAL WEAR. THE INSTRUCTIONS FOR USE STATES THAT RUPTURES CAN OCCUR AT ANY TIME AFTER IMPLANTATION, BUT IT IS MORE LIKELY TO OCCUR THE LONGER THE IMPLANT IS IMPLANTED. THE FOLLOWING MAY CAUSE IMPLANTS TO RUPTURE: STRESSING THE IMPLANT DURING IMPLANTATION OR OTHER PROCEDURES AND WEAKENING IT; FOLDING OR WRINKLING OF THE IMPLANT SHELL. BREAST IMPLANTS MAY ALSO WEAR OVER TIME. A SECOND PRODUCT WAS RECEIVED (LOT-5951355). NO ADVERSE EVENTS WERE REPORTED FOR THIS CONCOMITANT (CONTRALATERAL) DEVICE, THEREFORE NO FURTHER ANALYSIS IS REQUIRED. IT SHOULD BE NOTED THAT AS PART OF MENTOR¿S QUALITY PROCESS, ALL DEVICES ARE MANUFACTURED, INSPECTED, AND RELEASED TO APPROVED SPECIFICATIONS. ADDITIONAL COMPLAINT INFORMATION MONITORING FOR POTENTIAL SAFETY SIGNALS WILL BE CONDUCTED THROUGH COMPLAINT TRENDING AS PART OF POST-MARKET SURVEILLANCE. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).
SINCE THE DEVICE HAS NOT BEEN RETURNED FOR ANALYSIS, NO PRODUCT FAILURE ANALYSIS CAN BE CONDUCTED, AND NO DETERMINATION OF POSSIBLE CONTRIBUTING FACTORS CAN BE MADE. AS SUCH, THE INVESTIGATION WILL BE CLOSED. IF THE COMPLAINT DEVICE IS RECEIVED IN THE FUTURE, THE INVESTIGATION WILL BE REOPENED AND CONDUCTED AS APPROPRIATE. A MANUFACTURING RECORD EVALUATION (MRE) WAS PERFORMED, AND NO ANOMALIES WERE FOUND RELATED TO THIS COMPLAINT. IN ADDITION, THE MRE VERIFIES THAT THE DEVICE WAS MANUFACTURED IN ACCORDANCE WITH DOCUMENTED SPECIFICATION AND PROCEDURES. REASON FOR DEVICE EXPLANT AND/OR REOPERATION: RUPTURE. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).
IT WAS REPORTED THAT A (B)(6) YEAR-OLD FEMALE PATIENT UNDERWENT A BREAST AUGMENTATION REVISION WITH A 375CC MENTOR MEMORYGEL BREAST IMPLANT AND EXPERIENCED A RUPTURE ON THE LEFT SIDE POST-OPERATIVELY. AS A RESULT, THE PATIENT UNDERWENT EXPLANTATION ON (B)(6) 2021.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1204804 | MENTOR MEMORYGEL BREAST IMPLANT | PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED | FTR | MENTOR TEXAS | 3503751BC | 5947743 | 00081317000310 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 54 YR | Required Intervention |