FDA Adverse Event
Malfunction
Summary report: N
WARM TOUCH BLOWER MODEL 5800
MDR report key: 1230866
·
Received November 6, 2008
Report
- Report Number
- 2936999-2008-00558
- Event Type
- Malfunction
- Date Received
- November 6, 2008
- Date of Event
- September 1, 2008
- Report Date
- October 14, 2008
- Manufacturer
- COVIDIEN, FORMERLY TYCO HEALTHCARE
- Product Code
- DWJ
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
THE UNIT WAS EVALUATED AND THE BLOWER WAS FOUND TO BE DEFECTIVE; THE INSULATING RESISTANCE WAS LOW. THE CUSTOMER'S REPORTED FAILURE WAS CONFIRMED. THE BLOWER WAS REPLACED. THE DAMAGED KEYPAD WAS REPLACED. ALL FUNCTIONS PERFORMED AND PASSED BY THE ASSOCIATION OF FOREIGN ELECTRICIANS.
Description of Event or Problem · 1
CUSTOMER COMPLAINED OF ELECTRICAL SMELL AND INSULATION FAILURE. THE "OP" PERSON SWITCHED THE UNIT ON AND THEN THEY NOTED THE CABLE SMELL. THEY THEN TURNED THE UNIT OFF. THE UNIT WAS RUNNING TWO MINUTES AND, THEREFORE, THE CUSTOMER WAS UNABLE TO DETERMINE IF THE UNIT WOULD HAVE PERFORMED THE STEP-DOWN TO GO TO A LOWER TEMPERATURE OR TURN OFF ON IT'S OWN.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | WARM TOUCH BLOWER MODEL 5800 | CONVECTIVE AIR WARMER | DWJ | COVIDIEN, FORMERLY TYCO HEALTHCARE |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |