FDA Adverse Event Malfunction Summary report: N

WARM TOUCH BLOWER MODEL 5800

MDR report key: 1230866 · Received November 6, 2008

Report

Report Number
2936999-2008-00558
Event Type
Malfunction
Date Received
November 6, 2008
Date of Event
September 1, 2008
Report Date
October 14, 2008
Manufacturer
COVIDIEN, FORMERLY TYCO HEALTHCARE
Product Code
DWJ
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THE UNIT WAS EVALUATED AND THE BLOWER WAS FOUND TO BE DEFECTIVE; THE INSULATING RESISTANCE WAS LOW. THE CUSTOMER'S REPORTED FAILURE WAS CONFIRMED. THE BLOWER WAS REPLACED. THE DAMAGED KEYPAD WAS REPLACED. ALL FUNCTIONS PERFORMED AND PASSED BY THE ASSOCIATION OF FOREIGN ELECTRICIANS.

Description of Event or Problem · 1

CUSTOMER COMPLAINED OF ELECTRICAL SMELL AND INSULATION FAILURE. THE "OP" PERSON SWITCHED THE UNIT ON AND THEN THEY NOTED THE CABLE SMELL. THEY THEN TURNED THE UNIT OFF. THE UNIT WAS RUNNING TWO MINUTES AND, THEREFORE, THE CUSTOMER WAS UNABLE TO DETERMINE IF THE UNIT WOULD HAVE PERFORMED THE STEP-DOWN TO GO TO A LOWER TEMPERATURE OR TURN OFF ON IT'S OWN.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 WARM TOUCH BLOWER MODEL 5800 CONVECTIVE AIR WARMER DWJ COVIDIEN, FORMERLY TYCO HEALTHCARE

Patients

Seq Age Sex Outcome Treatment
1