BD RAPID DETECTION OF SARS-COV-2 VERITOR
Report
- Report Number
- 1119779-2021-01349
- Event Type
- Malfunction
- Date Received
- August 11, 2021
- Date of Event
- August 5, 2021
- Report Date
- September 21, 2021
- Manufacturer
- BECTON, DICKINSON & CO. (SPARKS)
- Product Code
- QKP
- PMA / PMN Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SN
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
INVESTIGATION SUMMARY: THIS STATEMENT IS TO SUMMARIZE THE INVESTIGATION RESULTS REGARDING THE COMPLAINT THAT ALLEGES FALSE POSITIVE RESULTS WHEN USING KIT RAPID DETECTION OF SARS-COV-2 VERITOR CE (MATERIAL # 256089), BATCH NUMBER 1181838. BD QUALITY PERFORMS A SYSTEMATIC APPROACH TO INVESTIGATE FALSE POSITIVE COMPLAINTS. THIS APPROACH INVOLVES REVIEW OF MANUFACTURING BATCH HISTORY RECORDS, TESTING OF RETENTION SAMPLES, AND TESTING OF CUSTOMER RETURNED SAMPLES, IF APPLICABLE. AN INVESTIGATION AND TESTING WERE PERFORMED ON THE BATCH NUMBER PROVIDED. A PHOTO WAS PROVIDED AND SHOWED ¿COV2: +¿ IN THE DISPLAY SCREEN. THE REPORTED WAS UNABLE TO BE CONFIRMED. THE ROOT CAUSE COULD NOT BE IDENTIFIED. BD QUALITY WILL CONTINUE TO CLOSELY MONITOR FOR TRENDS. H3 OTHER TEXT : SEE H10.
IT WAS REPORTED WHILE TESTING FOR SARS COV-2 A FALSE POSITIVE RESULT WAS OBTAINED. A REPEAT TEST WAS PERFORMED USING PCR AND THE RESULT WAS NEGATIVE. THERE WAS NO INDICATION THAT RESULTS WERE REPORTED OUT AND THERE WAS NO REPORT OF PATIENT IMPACT. EUA#: (B)(4).
MEDICAL DEVICE EXPIRATION DATE: UNKNOWN. DEVICE MANUFACTURE DATE: UNKNOWN. A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED, BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED.
IT WAS REPORTED WHILE TESTING FOR SARS COV-2 A FALSE POSITIVE RESULT WAS OBTAINED. A REPEAT TEST WAS PERFORMED USING PCR AND THE RESULT WAS NEGATIVE. THERE WAS NO INDICATION THAT RESULTS WERE REPORTED OUT AND THERE WAS NO REPORT OF PATIENT IMPACT. EUA#: (B)(4). THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: " REPORTING COMPLAINT AS FALSE POSITIVE RESULT."
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1204451 | BD RAPID DETECTION OF SARS-COV-2 VERITOR | CORONAVIRUS ANTIGEN DETECTION SYSTEM | QKP | BECTON, DICKINSON & CO. (SPARKS) | UNKNOWN |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |