NEOPUFF INFANT RESUSCITATOR
Report
- Report Number
- 9611451-2008-00631
- Event Type
- Malfunction
- Date Received
- October 28, 2008
- Report Date
- March 27, 2008
- Manufacturer
- FISHER & PAYKEL HEALTHCARE, LTD.
- Product Code
- BTL
- PMA / PMN Number
- K892885
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- EI
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
THE ACTUAL COMPLAINT DEVICE WAS VISUALLY INSPECTED FOR A BROKEN GAS INLET PORT. RESULTS: THE GAS INLET PORT OF THE NEOPUFF WAS BROKEN AND SNAPPED OFF FROM THE PANEL. STRESS MARK WAS VISIBLE AROUND THE CUT OFF AREA. IT WAS NOTICED THAT THE GAS SUPPLY LINE ADAPTOR WAS FULLY INSERTED INTO THE GAS INLET PORT CAUSING A TIGHT CONNECTION. THE SNAPPED OFF INLET PORT WAS STILL CONNECTED TO THE BROKEN GAS SUPPLY LINE WHEN THE NEOPUFF WAS INVESTIGATED. CONCLUSION: BECAUSE OF THE TIGHT CONNECTION, IT APPEARS THAT AN EXTRA AMOUNT OF FORCE WAS APPLIED TO DISCONNECT THE GAS SUPPLY LINE FROM THE GAS INLET PORT RESULTING IN SNAPPING OFF THE INLET PORT. THE PORTS ON THE NEOPUFF USE A TAPERED FIT TO HOLD THE CONNECTING TUBES IN PLACE. REMOVAL OF THE GAS SUPPLY LINE FROM THE GAS INLET PORT OF THE NEOPUFF IS NORMALLY ACHIEVED BY CLASPING AND PULLING THE PLASTIC CONNECTOR AWAY FROM THE INLET PORT. IF THE GAS SUPPLY LINE IS FULLY INSERTED INTO THE GAS INLET PORT AND REMOVAL IS ATTEMPTED BY WAGGLING OR PULLING IT SIDEWAYS, IT COULD LEAD TO CRACKING OR SNAPPING OFF OF THE GAS INLET PORT AFTER MULTIPLE REUSES OF THE NEOPUFF.
A HOSPITAL REPORTED THAT THE GAS INLET PORT OF AN RD900AEU NEOPUFF INFANT RESUSCITATOR WAS BROKEN OFF. NO PT CONSEQUENCE WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | NEOPUFF INFANT RESUSCITATOR | BTL | FISHER & PAYKEL HEALTHCARE, LTD. | RD900AEU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |