FLO-GARD 6201 VOLUMETRIC INFUSION PUMP
Report
- Report Number
- 6000001-2007-88786
- Event Type
- Malfunction
- Date Received
- November 17, 2008
- Date of Event
- June 19, 2007
- Report Date
- June 19, 2007
- Manufacturer
- BAXTER HEALTHCARE
- Product Code
- FRN
- PMA / PMN Number
- K915522
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC, US
- Reporter Occupation
- PHARMACIST
Narratives
THE FACILITY REPRESENTATIVE REPORTED A BROKEN DOOR. INFORMATION WAS NOT PROVIDED REGARDING WHETHER OR NOT THE PROBLEM OCCURRED DURING A PATIENT INFUSION.EVALUATION SUMMARY: THE PUMP WAS EVALUATED BY A BAXTER SERVICE TECHNICIAN. THE REPORTED CONDITION "DOOR NEEDS REPAIR" WAS CONFIRMED CAUSED BY A CRACKED DOOR AT HINGES. THE DOOR ASSEMBLY WAS REPLACED. REVIEW OF THE COMPLAINT HISTORY REVEALS SIMILAR REPORTS HAVE BEEN RECEIVED FOR THIS PRODUCT FOR THE REPORTED ISSUE. THIS ISSUE IS BEING INVESTIGATED UNDER CAPA.
THE DEVICE WAS RETURNED FOR SERVICE. DURING SERVICE BY BAXTER, A CRACKED DOOR AT HINGES WAS FOUND. ACCORDING TO THE FACILITY REPRESENTATIVE, NO PATIENT INJURY OR MEDICAL INTERVENTION HAD BEEN REPORTED RELATED TO THE DEVICE. NO ADDITIONAL INFORMATION OR CONTACT WAS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | FLO-GARD 6201 VOLUMETRIC INFUSION PUMP | 80FRN | FRN | BAXTER HEALTHCARE | N/A | N/A |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |