DA VINCI SURGICAL SYSTEM
Report
- Report Number
- 2955842-2008-01360
- Event Type
- Other
- Date Received
- November 7, 2008
- Date of Event
- October 17, 2008
- Report Date
- November 7, 2008
- Manufacturer
- INTUITIVE SURGICAL, INC.
- Product Code
- NAY
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- OTHER
Narratives
CONCLUSIONS- THE INVESTIGATION CONDUCTED BY FIELD SERVICE ENGINEERING WAS UNABLE TO CONFIRM THE UNRESPONSIVE PSM ISSUE REPORTED BY THE CUSTOMER. SYSTEM ERROR CODE #21003 IS ASSOCIATED WITH A PT SIDE MANIPULATOR (PSM) ARM. THE PSM IS AN INSTRUMENT ARM LOCATED ON THE PT SIDE CART, THAT PROVIDES THE STERILE INTERFACE FOR THE ENDOWRIST INSTRUMENTS. THE SYSTEM ALARM (SYSTEM GENERATED FAULT CODE) FUNCTIONED AS DESIGNED AND THERE WAS NO INJURY TO THE PT. THE #21003 SYSTEM ERROR CODE APPEARS WHEN THE DA VINCI SAFETY SYSTEM DETECTS THAT A SYSTEM ARM WAS "BUMPED" OR RESTRICTED FROM MOVING PROPERLY DURING STARTUP. UPON DETERMINING THESE CONDITIONS, THE SAFETY SYSTEMS PUT DA VINCI IN A "RECOVERABLE SAFE STATE". AS OF (B) (6), 2008, THERE HAVE BEEN NO REPORTED RECURRENCES OF THE ISSUE AT THIS HOSPITAL.
IT WAS REPORTED THAT DURING A DA VINCI HYSTERECTOMY PROCEDURE ONE OF THE PT'S SIDE MANIPULATOR (PSM) ARMS WOULD NOT RESPOND AND THE FAULT OVER-RIDE LIGHT WAS ILLUMINATED. WITH THE ASSISTANCE OF AN ISI TECHNICAL SUPPORT ENGINEER THE SYSTEM ERROR LOGS WERE RETRIEVED AND SYSTEM ERROR #21003 WAS FOUND. THE CUSTOMER TRIED TO ENGAGE AN INSTRUMENT HOWEVER, THE ENGAGEMENT ISSUE CONTINUED TO OCCUR. THE SURGEON DECIDED TO CONVERT TO TRADITIONAL OPEN SURGICAL TECHNIQUES TO COMPLETE THE PLANNED PROCEDURE, NO PT HARM WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | DA VINCI SURGICAL SYSTEM | ENDOSCOPIC INSTRUMENT CONTROL SYSTEM | NAY | INTUITIVE SURGICAL, INC. | IS1200 A4. 3P9 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other | DA VINCI SURGICAL SYSTEM| INSTRUMENTS| ACCESSORIES. |