FDA Adverse Event Other Summary report: N

DA VINCI S SURGICAL SYSTEM

MDR report key: 1230292 · Received November 7, 2008

Report

Report Number
2955842-2008-01359
Event Type
Other
Date Received
November 7, 2008
Date of Event
October 17, 2008
Report Date
November 7, 2008
Manufacturer
INTUITIVE SURGICAL, INC.
Product Code
NAY
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OK, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

RESULT AND CONCLUSIONS- AN INVESTIGATION CONDUCTED BY THE FIELD SERVICE ENGINEER CONCLUDED THE VISION ISSUES EXPERIENCED BY THE CUSTOMER WERE ASSOCIATED WITH THE HIGH RESOLUTION STEREO VIEWER (HRSV). THE HIGH RESOLUTION STEREO VIEWER (HRSV) IS LOCATED ON THE SURGEON CONSOLE AND PROVIDES THE VIDEO IMAGE FOR THE SURGEON CONSOLE OPERATOR. THE SYSTEM WAS REPAIRED BY REPLACING THE AFFECTED HRSV. AS OF 2008, THERE HAVE BEEN NO REPORTED RECURRENCES OF THE ISSUE AT THIS HOSPITAL.

Description of Event or Problem · 1

IT WAS REPORTED THAT PRIOR TO STARTING A DA VINCI S PROSTATECTOMY PROCEDURE, THE RIGHT EYE IN THE HIGH RESOLUTION STEREO VIEWER (HRSV) WAS DISPLAYING HORIZONTAL LINES. WITH THE ASSISTANCE OF AN INTUITIVE SURGICAL TECHNICAL SUPPORT REPRESENTATIVE, THE SITE RESET THE SURGEON SIDE CONSOLE, RESEATED THE VISION CABLE, AND CYCLED POWER, HOWEVER, THE PROBLEM WAS NOT RESOLVED. THE PT WAS UNDER ANESTHESIA AND PORT INCISIONS HAD BEEN CREATED WHEN THE SURGICAL STAFF MADE THE DECISION TO ABORT THE PROCEDURE. NO PT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 DA VINCI S SURGICAL SYSTEM ENDOSCOPIC INSTRUMENT CONTROL SYSTEM NAY INTUITIVE SURGICAL, INC. IS2000 A5.1P6

Patients

Seq Age Sex Outcome Treatment
1 Other DA VINCI S SURGICAL SYSTEM INSTRUMENTS| ACCESSORIES.