COLLEAGUE CX TRIPLE CHANNEL VOLUMETRIC INFUSION PUMP
Report
- Report Number
- 6000001-2007-88576
- Event Type
- Malfunction
- Date Received
- November 17, 2008
- Date of Event
- June 21, 2007
- Report Date
- June 21, 2007
- Manufacturer
- BAXTER HEALTHCARE PTE. LTD.
- Product Code
- FRN
- PMA / PMN Number
- K010566
- Removal / Correction Number
- 6000001-06/21/07-004-R
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- OTHER
Narratives
EVALUATION SUMMARY: THIS REPORT IS BEING RESUBMITTED IN ACCORDANCE WITH INSTRUCTIONS FROM FDA TO ADDRESS A MANUFACTURER REPORT SEQUENCE NUMBER ISSUE THAT OCCURRED WHEN THIS MDR OR SUPPLEMENT WAS ORIGINALLY SUBMITTED TO THE FDA ON JUN 26 2007. THE DEVICE HAS BEEN REQUESTED BY BAXTER QUALITY MANAGEMENT FOR EVALUATION BUT HAS NOT YET BEEN RECEIVED. SHOULD THE PUMP BE RECEIVED FOR EVALUATION, A FOLLOW UP REPORT WILL BE FILED UPON COMPLETION OF THE EVALUATION OR IF ADDITIONAL INFORMATION BECOMES AVAILABLE.THE 16:310:867:0002 IS DUE TO A SOFTWARE "MAILBOX", THAT STORES TEMPORARY DATA UNTIL IT CAN BE PROCESSED, BECOMING FILLED. THE ISSUE IDENTIFIED ABOVE HAS BEEN REPRODUCED IN OUR FACILITY WITH USER INTERFACE SOFTWARE VERSIONS 5.09.90 AND 6.13.90. PRELIMINARY ANALYSIS INDICATES THAT THIS ISSUE OCCURS ON TRIPLE CHANNEL DEVICES, DURING USER PROGRAMMING, WITH THREE CHANNELS INFUSING, IN SPECIFIC USE CASE SCENARIOS. ANALYSIS OF FIELD INFORMATION AND THE DATA, EVENT LOGS AND THE ISSUE INVESTIGATION HAVE NOT SHOWN THAT THE ISSUE WILL OCCUR ON PREVIOUS SOFTWARE VERSIONS IN THE SAME USE CASE SCENARIO. THERE IS NO EVIDENCE THAT THIS CAN OCCUR ON SINGLE CHANNEL DEVICES. ROOT CAUSE INVESTIGATIONS ARE CONTINUING. REVIEW OF THE COMPLAINT HISTORY REVEALS SIMILAR REPORTS HAVE BEEN RECEIVED FOR THIS PRODUCT FOR THE REPORTED ISSUE. THIS ISSUE IS BEING INVESTIGATED UNDER CAPA.
A BAXTER REPRESENTATIVE REPORTED THAT THE FACILITY'S BIOMEDICAL DIRECTOR REPORTED AN INCIDENT OF FAILURE CODE 16:310:867:002 ON A COLLEAGUE CX TRIPLE CHANNEL INFUSION PUMP RECENTLY UPGRADED. THE EVENT REPORTEDLY OCCURRED DURING SET UP. NO REPORT OF ANY PATIENT INJURY. NO ADDITIONAL INFORMATION WAS PROVIDED. THE PUMP EVENT HISTORY WAS PROVIDED WITH THE REPORT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | COLLEAGUE CX TRIPLE CHANNEL VOLUMETRIC INFUSION PUMP | 80FRN | FRN | BAXTER HEALTHCARE PTE. LTD. | N/A |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |