ANEURX AAADVANTAGE STENT GRAFT SYSTEM - HYDRO
Report
- Report Number
- 2953200-2008-01043
- Event Type
- Injury
- Date Received
- November 13, 2008
- Date of Event
- October 21, 2008
- Report Date
- October 21, 2008
- Manufacturer
- MEDTRONIC CARDIOVASCULAR
- Product Code
- MIH
- PMA / PMN Number
- P990020
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- PHYSICIAN
Narratives
EVALUATION: RESULTS: SEVERE CALCIFICATION IN THE AORTIC NECK), (ENDOLEAK). CONCLUSION: (SEVERE CALCIFICATION IN THE AORTIC NECK). SECONDARY INTERVENTION.
A TALENT ABDOMINAL STENT GRAFT SYSTEM WAS IMPLANTED IN A PT FOR THE ENDOVASCULAR TREATMENT OF AN ABDOMINAL AORTIC ANEURYSM. THE AORTIC NECK WAS SEVERELY CALCIFIED. IT WAS REPORTED THE BIFURCATED STENT GRAFT WAS IMPLANTED RIGHT UNDER THE RENAL ARTERY. THE AORTIC NECK WAS 26 MM IN DIAMETER. THE ANGIOGRAM REVEALED A TYPE I ENDOLEAK WHICH WAS DUE TO THE SEVERE CALCIFICATION IN THE AORTIC NECK. THE PHYSICIAN BALLOONED THREE TIMES WITH A RELIANT BALLOON, THE TYPE I ENDOLEAK IMPROVED WITH EACH INFLATION, HOWEVER, THE ENDOLEAK DID NOT COMPLETELY RESOLVE. THE PHYSICIAN PLACED ANOTHER MANUFACTURER'S STENT AND THE ENDOLEAK RESOLVED. NO ADDITIONAL CLINICAL SEQUELAE WERE REPORTED AND THE PT IS FINE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ANEURX AAADVANTAGE STENT GRAFT SYSTEM - HYDRO | MIH | MEDTRONIC CARDIOVASCULAR | NA | V00117275 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 69 YR | Required Intervention |