FDA Adverse Event Injury Summary report: N

ANEURX AAADVANTAGE STENT GRAFT SYSTEM - HYDRO

MDR report key: 1230190 · Received November 13, 2008

Report

Report Number
2953200-2008-01043
Event Type
Injury
Date Received
November 13, 2008
Date of Event
October 21, 2008
Report Date
October 21, 2008
Manufacturer
MEDTRONIC CARDIOVASCULAR
Product Code
MIH
PMA / PMN Number
P990020
Removal / Correction Number
NA
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

EVALUATION: RESULTS: SEVERE CALCIFICATION IN THE AORTIC NECK), (ENDOLEAK). CONCLUSION: (SEVERE CALCIFICATION IN THE AORTIC NECK). SECONDARY INTERVENTION.

Description of Event or Problem · 1

A TALENT ABDOMINAL STENT GRAFT SYSTEM WAS IMPLANTED IN A PT FOR THE ENDOVASCULAR TREATMENT OF AN ABDOMINAL AORTIC ANEURYSM. THE AORTIC NECK WAS SEVERELY CALCIFIED. IT WAS REPORTED THE BIFURCATED STENT GRAFT WAS IMPLANTED RIGHT UNDER THE RENAL ARTERY. THE AORTIC NECK WAS 26 MM IN DIAMETER. THE ANGIOGRAM REVEALED A TYPE I ENDOLEAK WHICH WAS DUE TO THE SEVERE CALCIFICATION IN THE AORTIC NECK. THE PHYSICIAN BALLOONED THREE TIMES WITH A RELIANT BALLOON, THE TYPE I ENDOLEAK IMPROVED WITH EACH INFLATION, HOWEVER, THE ENDOLEAK DID NOT COMPLETELY RESOLVE. THE PHYSICIAN PLACED ANOTHER MANUFACTURER'S STENT AND THE ENDOLEAK RESOLVED. NO ADDITIONAL CLINICAL SEQUELAE WERE REPORTED AND THE PT IS FINE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ANEURX AAADVANTAGE STENT GRAFT SYSTEM - HYDRO MIH MEDTRONIC CARDIOVASCULAR NA V00117275

Patients

Seq Age Sex Outcome Treatment
1 69 YR Required Intervention