FDA Adverse Event Malfunction Summary report: N

PUMP MMT-1782K 670G V4.11 MM

MDR report key: 12299804 · Received August 10, 2021

Report

Report Number
2032227-2021-178745
Event Type
Malfunction
Date Received
August 10, 2021
Date of Event
July 30, 2021
Report Date
December 16, 2021
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OZP
UDI-DI
000000763000187422
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

CUSTOMER COMPLAINED ON (B)(6) 2021 INSULIN PUMP ALARMED STUCK BUTTON. DEVICE PASSED SELF TEST AND DISPLACEMENT TEST. ALL BUTTONS FUNCTION PROPERLY. DEVICE SUCCESSFULLY DOWNLOADED TO THUS. NO STUCK BUTTON ERROR ALARMS NOTED DURING TESTING. VERIFIED DEVICE ALARMED STUCK BUTTON ON (B)(6) 202112:14:25.000 IN DEVICE DOWNLOADED HISTORY. DEVICE PASSED THE KEYPAD VOLTAGE TEST. NO DAMAGE WAS NOTED TO KEYPAD ASSEMBLY AND CONNECTOR ON ELECTRICAL BOARD 1 WAS LOCKED PROPERLY DURING VISUAL INSPECTION. THE ELECTRONIC ASSEMBLIES WERE INSPECTED AND NO ANOMALIES NOTED. THE FOLLOWING WERE NOTED DURING VISUAL INSPECTION: CORRODED BATTERY TUBE, SCRATCHED CASE, PILLOWING KEYPAD OVERLAY, SERIAL NUMBER LABEL MISSING AND END CAP ADDRESS LABEL. THE P-CAP/RESERVOIR DOES LOCK PROPERLY. ALL BUTTONS FUNCTIONED PROPERLY DURING TEST. NO STUCK BUTTON ERROR ALARMS NOTED DURING TESTING. VERIFIED DEVICE ALARMED STUCK BUTTON ON (B)(6) 2021 12:14:25.000 IN DEVICE DOWNLOADED HISTORY. MEDTRONIC, INC. (MEDTRONIC) IS SUBMITTING THIS REPORT TO COMPLY WITH 21 C.F.R. PART 803, THE MEDICAL DEVICE REPORTING REGULATION. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION IN THE TIME ALLOTTED AND HAS PROVIDED AS MUCH INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY "DEFECTS" OR HAS "MALFUNCTIONED". THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA, TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REPORTING PURSUANT TO PART 803. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE IT BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT.

Additional Manufacturer Narrative · 1

(B)(4). CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Description of Event or Problem · 1

INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD A KEYPAD ANOMALY. CUSTOMER STATED THEY HAD STUCK BUTTON ALARM. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1202288 PUMP MMT-1782K 670G V4.11 MM ARTIFICIAL PANCREAS DEVICE SYSTEM, SINGLE HORMONAL CONTROL OZP MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-1782K HG384AR 000000763000187422

Patients

Seq Age Sex Outcome Treatment
1 Unknown