FDA Adverse Event Death Summary report: N

WHEELCHAIR SAFETY BELT

MDR report key: 12299 · Received February 1, 1994

Report

Report Number
12299
Event Type
Death
Date Received
February 1, 1994
Date of Event
January 17, 1994
Report Date
January 28, 1994
Manufacturer
ALI MED
Product Code
IQB
Adverse Event
Yes
Report Source
User Facility report
Reporter Location
MN, US
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 1

RESIDENT WAS SITTING IN WHEELCHAIR AT RNS DESK. SHE WAS SEEN TO LEAN FORWARD AND FELL OUT OF WHEELCHAIR, LANDING ON RIGHT SIDE ON FLOOR. SAFETY SEAT BELT HAD BEEN CLOSED ACCORDING TO NURSING ASSISTANTS CARING FOR RESIDENT. SEAT BELT WAS NOT CLOSED AFTER FALL. PT HAD SKIN TEAR BELOW RIGHT KNEE. SHE HIT HEAD AGAINST RNS DESK. NEURO CHECKS WERE DONE FOR 3 DAYS. PHYSICIAN NOTIFIED OF FALL AND POSSIBLE INJURY-PROTRUSION OF RIGHT KNEE AREA. NO IMMEDIATE INTERVENTION ORDERED. PHYSICIAN VISIT ON 1/19//94 AND X-RAY ORDERED. EXAM OF SAFETY BELT AFTER FALL REVEALED THAT ONLY MINIMAL PRESSURE ON RELEASE BUTTON WAS NEEDED TO RELEASE THE SAFETY BELT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 WHEELCHAIR SAFETY BELT IQB ALI MED

Patients

Seq Age Sex Outcome Treatment
1 98 YR Not Applicable| D