BD INTERLINK BLUNT PLASTIC CANNULA
Report
- Report Number
- 1911916-2021-00801
- Event Type
- Malfunction
- Date Received
- August 9, 2021
- Date of Event
- July 12, 2021
- Report Date
- August 3, 2021
- Manufacturer
- BD MEDICAL (BD WEST) MEDICAL SURGICAL
- Product Code
- FMI
- UDI-DI
- 30382903033455
- PMA / PMN Number
- K974363
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OR, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
DATE OF EVENT: UNKNOWN. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED. THERE WERE MULTIPLE LOT NUMBERS REPORTED TO BE INVOLVED. THE INFORMATION FOR EACH LOT NUMBER IS AS FOLLOWS: MEDICAL DEVICE LOT #: 1015092, MEDICAL DEVICE EXPIRATION DATE: 2026-03-31, DEVICE MANUFACTURE DATE: 2021-01-15. MEDICAL DEVICE LOT #: 8090585, MEDICAL DEVICE EXPIRATION DATE: 2023-05-31, DEVICE MANUFACTURE DATE: 2018-03-31. MEDICAL DEVICE LOT #: 1096756, MEDICAL DEVICE EXPIRATION DATE: 2026-04-30, DEVICE MANUFACTURE DATE: 2021-04-06. INVESTIGATION SUMMARY: IT WAS REPORTED BY THE DISTRIBUTOR THAT PIECES OF RUBBER FROM THE VIAL ARE BEING SHEARED OFF BY THE CANNULA AND DRAWN UP. TO AID IN THE INVESTIGATION, ONE PHOTO WAS PROVIDED FOR EVALUATION BY OUR QUALITY TEAM. THE PHOTO SHOWS A VIAL OR PROPOFOL. FROM THE PHOTO THERE APPEARS TO BE A SMALL PIECE OF FOREIGN MATTER INSIDE THE VIAL. IT COULD BE POSSIBLE THIS PRODUCT IS NOT BEING USED AS INTENDED. THE BLUNT PLASTIC CANNULA IS NOT DESIGNED TO BE USED WITH STOPPER VIALS. A DEVICE HISTORY RECORD REVIEW WAS COMPLETED FOR PROVIDED MATERIAL NUMBER 303345, LOT NUMBERS 1015092, 8090585 AND 1096756. THE REVIEW DID NOT REVEAL ANY DETECTED QUALITY ISSUES DURING THE PRODUCTION OF THESE LOTS THAT COULD HAVE CONTRIBUTED TO THE REPORTED DEFECT. BASED ON THE INVESTIGATION AND WITH THE PHOTO SAMPLE ANALYSIS THE SYMPTOM REPORTED BY THE CUSTOMER IS CONFIRMED. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. OUR QUALITY TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS.
IT WAS REPORTED THAT 2 BD INTERLINK¿ BLUNT PLASTIC CANNULAS EACH FROM LOTS 1015092, 8090585, AND 1096756 SHEARED OFF THE RUBBER VIAL TOP INTO THE MEDICATION WHEN DRAWING IT UP. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "THE RUBBER FROM THE TOP OF THE MEDICATION VIAL IS BEING SHEARED OFF WITH THE CANNULA AND GOING INTO THE MEDICATION VIAL."
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1195315 | BD INTERLINK BLUNT PLASTIC CANNULA | HYPODERMIC SINGLE LUMEN NEEDLE | FMI | BD MEDICAL (BD WEST) MEDICAL SURGICAL | 303345 | SEE SECTION H.10. | 30382903033455 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |