FDA Adverse Event Malfunction Summary report: N

OMNIPOD INSULIN MANAGEMENT SYSTEM

MDR report key: 12295570 · Received August 9, 2021

Report

Report Number
3004464228-2021-13663
Event Type
Malfunction
Date Received
August 9, 2021
Date of Event
July 29, 2021
Report Date
July 29, 2021
Manufacturer
INSULET CORPORATION
Product Code
LZG
UDI-DI
20385081120033
PMA / PMN Number
K192659
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE HAS NOT BEEN RETURNED/RECEIVED TO DATE. IF THE DEVICE IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED WITH THE INVESTIGATION RESULTS. WE ARE UNABLE TO CONFIRM THE BENT CANNULA OR TO DETERMINE IF IT COULD HAVE CONTRIBUTED TO THE REPORTED HYPERGLYCEMIA. LOT RELEASE RECORDS WERE REVIEWED AND THE PRODUCT LOT MET ALL ACCEPTANCE CRITERIA. SPECIFICALLY, A POD IS PAIRED TO A PDM AND PUT THROUGH SIMULATED USE TESTING INCLUDING CONFIRMING THE BLUE SOFT CANNULA IS INSPECTED FOR ANY DAMAGE.

Description of Event or Problem · 1

IT WAS REPORTED THE PATIENT'S BLOOD GLUCOSE LEVEL ROSE TO 202, 259, 289, 290, AND 315 MG/DL WHILE WEARING THE POD BETWEEN 4 AND 24 HOURS. WHEN REMOVED FROM THE INFUSION SITE (HIP/BUTTOCKS), THE POD'S CANNULA WAS FOUND BENT. AS TREATMENT, A NEW POD WAS APPLIED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1192147 OMNIPOD INSULIN MANAGEMENT SYSTEM PUMP, INFUSION, INSULIN LZG INSULET CORPORATION 19191 L71214 20385081120033

Patients

Seq Age Sex Outcome Treatment
1 51 YR