FDA Adverse Event Malfunction Summary report: N

ULTRANET COLLIMATOR

MDR report key: 122943 · Received September 26, 1997

Report

Report Number
57331-1997-00032
Event Type
Malfunction
Date Received
September 26, 1997
Date of Event
August 12, 1997
Report Date
September 26, 1997
Manufacturer
GENERAL ELECTRIC CO.
Product Code
KPW
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 1

IT WAS REPORTED TO GE THAT THE COLLIMATOR DETACHED FROM ITS MOUNT AND FELL. APPARENTLY, THE MOUNTING SCREWS BACKED OUT ALLOWING IT TO FALL. THE SCREWS WERE REPLACED AND SECURED AND THE UNIT PLACED BACK IN SERVICE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ULTRANET COLLIMATOR DIAGNOSTIC X-RAY KPW GENERAL ELECTRIC CO. 45296017 NA

Patients

Seq Age Sex Outcome Treatment
1 *