FDA Adverse Event
Malfunction
Summary report: N
ULTRANET COLLIMATOR
MDR report key: 122943
·
Received September 26, 1997
Report
- Report Number
- 57331-1997-00032
- Event Type
- Malfunction
- Date Received
- September 26, 1997
- Date of Event
- August 12, 1997
- Report Date
- September 26, 1997
- Manufacturer
- GENERAL ELECTRIC CO.
- Product Code
- KPW
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- OTHER
Narratives
Description of Event or Problem · 1
IT WAS REPORTED TO GE THAT THE COLLIMATOR DETACHED FROM ITS MOUNT AND FELL. APPARENTLY, THE MOUNTING SCREWS BACKED OUT ALLOWING IT TO FALL. THE SCREWS WERE REPLACED AND SECURED AND THE UNIT PLACED BACK IN SERVICE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ULTRANET COLLIMATOR | DIAGNOSTIC X-RAY | KPW | GENERAL ELECTRIC CO. | 45296017 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | * |