FDA Adverse Event Malfunction Summary report: N

TRUE METRIX

MDR report key: 12293008 · Received August 9, 2021

Report

Report Number
1000113657-2021-00471
Event Type
Malfunction
Date Received
August 9, 2021
Date of Event
July 15, 2021
Report Date
October 5, 2021
Manufacturer
TRIVIDIA HEALTH INC
Product Code
NBW
UDI-DI
00021292007836
PMA / PMN Number
K140100
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

SECTIONS WITH ADDITIONAL INFORMATION AS OF 05-OCT-2021: H10: METER WAS NOT RETURNED FOR EVALUATION. TEST STRIPS WERE NOT RETURNED FOR EVALUATION. RETENTION TESTING WAS PERFORMED USING TEST STRIPS FROM THE SAME LOT. RETENTION STRIP LOT TESTED WITHIN SPECIFICATIONS. MOST LIKELY UNDERLYING ROOT CAUSE: MLC-062: USER HAD POOR TECHNIQUE.

Additional Manufacturer Narrative · 1

INTERNAL REPORT REFERENCE NUMBER: (B)(4). METER AND TEST STRIPS WERE NOT RETURNED FOR EVALUATION. NOTE: MANUFACTURER CONTACTED CUSTOMER IN A FOLLOW-UP CALL ON 28-JUL-2021 TO ENSURE THE REPLACEMENT PRODUCTS RESOLVED THE INITIAL CONCERN - ABLE TO ESTABLISH CONTACT WITH CUSTOMER WHO STATED REPLACEMENT PRODUCTS RESOLVED INITIAL CONCERN.

Description of Event or Problem · 1

CONSUMER REPORTED COMPLAINT FOR HIGH BLOOD GLUCOSE TEST RESULTS. CUSTOMER IS USING REPLACEMENT TRUE METRIX METER THAT WAS SENT ON INTERNAL REPORT REFERENCE NUMBER (B)(4). THE CUSTOMER IS CONCERNED WITH TEST RESULTS FROM RESULTS OBTAINED OF 223, 154 AND 202 MG/DL. THE CUSTOMER¿S EXPECTED AM FASTING BLOOD GLUCOSE TEST RESULT RANGE IS 80-90 MG/DL AND EXPECTED TWO HOUR POST MEAL EXPECTED RESULT IS 200 MG/DL. THE CUSTOMER FEELS WELL AND DID NOT REPORT ANY SYMPTOMS. MEDICAL ATTENTION IS NOT REPORTED AS A RESULT OF THE ACTUAL BLOOD GLUCOSE RESULTS. THE CUSTOMER WAS NOT USING THE PROPER TESTING TECHNIQUE. DURING THE CALL, A BLOOD TEST WAS PERFORMED BY THE CUSTOMER FASTING AND PRODUCED TEST RESULT OF 207 MG/DL USING TRUE METRIX METER; CUSTOMER WAS NOT SATISFIED WITH THE RESULT OBTAINED. THE PRODUCT IS STORED ACCORDING TO SPECIFICATION IN THE BEDROOM. THE TEST STRIP LOT MANUFACTURER¿S EXPIRATION DATE IS 06/24/2022 AND TEST STRIPS WERE OPENED ONE WEEK PRIOR TO CALL. THE METER MEMORY WAS REVIEWED FOR PREVIOUS TEST RESULT HISTORY (CUSTOMER WAS NOT CONCERNED WITH THE RESULT OF 66 MG/DL): (B)(6).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1192343 TRUE METRIX SYSTEM, TEST BLOOD GLUCOSE, OVER THE COUNTER NBW TRIVIDIA HEALTH INC STRIP, TRUE METRIX 100CTMG/DL ZX4202S 00021292007836

Patients

Seq Age Sex Outcome Treatment
1