FDA Adverse Event Injury Summary report: N

INSPIRA TEXTURED SILICONE GEL FILLED BREAST IMPLANT

MDR report key: 12291023 · Received August 6, 2021

Report

Report Number
9617229-2021-48232
Event Type
Injury
Date Received
August 6, 2021
Date of Event
November 15, 2020
Report Date
September 30, 2021
Manufacturer
ALLERGAN (COSTA RICA)
Product Code
FTR
PMA / PMN Number
P020056
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
HU
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 0

DEVICE EVALUATION: VISUAL ANALYSIS OF THE RETURNED DEVICE IDENTIFIED ONE EXTENDED OPENING, RED PARTICLES MATERIAL WAS FOUND ON THE SHELL AND FOLD CREASES. A WEIGHT TEST OF THE DEVICE WAS VERIFIED AND THE DEVICE UNDERWEIGHT. A MICROSCOPIC ANALYSIS WAS PERFORMED WHICH IDENTIFIED ONE OPENING SHARP AND WEAR ABRASION. A DIMENSION MEASUREMENT IN THE SHELL WAS PERFORMED WHICH IDENTIFY THE THICKNESS WITHIN SPECIFICATION. BASED ON THE DEVICE ANALYSIS THE FINAL ASSESSMENT IS: ONE SHARP OPENING IN THE SIDE RADIUS ASSESSED AS UNIDENTIFIED (TEAR) OPENING.

Description of Event or Problem · 0

PHYSICIAN REPORTED A RIGHT SIDE "IMPLANT RUPTURE" DIAGNOSED BY MRI. THE DEVICE HAS BEEN EXPLANTED.

Additional Manufacturer Narrative · 1

INITIAL REPORTER NAME AND ADDRESS: (B)(6). VISUAL ANALYSIS: VISUAL ANALYSIS OF THE PHOTOGRAPHS IDENTIFIED. DEVICE PATCH WITH LOT (OR CATALOG) NUMBER LOT NUMBER: 2032075. RED PARTICLES PARTICLE MATERIAL ON THE SHELL. OPENING EXTENDED ON RADIUS. A REVIEW OF THE DEVICE HISTORY RECORD HAS BEEN COMPLETED. NO DEVIATIONS OR NON-CONFORMANCES NOTED. FURTHER INFORMATION FROM THE REPORTER REGARDING EVENT, PRODUCT, OR PATIENT DETAILS HAS BEEN REQUESTED. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME. REASON FOR REOPERATION: RUPTURE.

Description of Event or Problem · 1

PHYSICIAN REPORTED "IMPLANT RUPTURE". DEVICE HAS BEEN EXPLANTED. THIS RELATES TO THE RIGHT SIDE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1183422 INSPIRA TEXTURED SILICONE GEL FILLED BREAST IMPLANT PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED FTR ALLERGAN (COSTA RICA) 2032075

Patients

Seq Age Sex Outcome Treatment
1 36 YR Required Intervention