0.5 ML BD ULTRA-FINE II INSULIN SYRINGE
Report
- Report Number
- 1920898-2021-00865
- Event Type
- Malfunction
- Date Received
- August 5, 2021
- Date of Event
- July 8, 2021
- Report Date
- August 16, 2021
- Manufacturer
- BD MEDICAL - DIABETES CARE
- Product Code
- FMF
- PMA / PMN Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TH
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
H6: INVESTIGATION SUMMARY: CUSTOMER RETURNED A NUMBER OF IMAGES SHOWING A NUMBER OF SHELF CARTONS FOR 0.5ML, 30 GAUGE, 8MM SYRINGES FROM LOT 0041321. ONE OF THE SHELF CARTONS IS MISSING THE PRINT FOR ITS LOT NUMBER, MANUFACTURING DATE, AND EXPIRATION DATE. A REVIEW OF THE DEVICE HISTORY RECORD WAS COMPLETED FOR BATCH# 0041321. ALL INSPECTIONS AND CHALLENGES WERE PERFORMED PER THE APPLICABLE OPERATIONS QC SPECIFICATIONS. THERE WERE ZERO (0) NOTIFICATIONS NOTED THAT PERTAINED TO THE COMPLAINT. BASED ON THE IMAGES RECEIVED, BD WAS ABLE TO CONFIRM THE CUSTOMER¿S INDICATED FAILURE OF LABEL INFORMATION MISSING. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. INFORMATION WILL BE CAPTURED ON TREND REPORTS AND MONITORED MONTHLY. OUR BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS. H3 OTHER TEXT : SEE H10.
IT WAS REPORTED THAT 2 0.5 ML BD ULTRA-FINE¿ II INSULIN SYRINGES EXPERIENCED MISSING LABEL INFORMATION. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: LOT NOT MATCH POSITION.
(B)(6). A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.
IT WAS REPORTED THAT 2 0.5 ML BD ULTRA-FINE" II INSULIN SYRINGES EXPERIENCED MISSING LABEL INFORMATION. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: LOT NOT MATCH POSITION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1176747 | 0.5 ML BD ULTRA-FINE II INSULIN SYRINGE | INSULIN SYRINGE | FMF | BD MEDICAL - DIABETES CARE | 0041321 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |