FDA Adverse Event Injury Summary report: N

ID NOW COVID 19 ASSAY

MDR report key: 12280775 · Received August 5, 2021

Report

Report Number
1221359-2021-02191
Event Type
Injury
Date Received
August 5, 2021
Date of Event
July 8, 2021
Report Date
November 23, 2021
Manufacturer
ABBOTT DIAGNOSTICS SCARBOROUGH, INC.
Product Code
QJR
UDI-DI
10811877011320
PMA / PMN Number
EUA2000074
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
BE
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

ADDITIONAL DATA: H10 TESTING WAS PERFORMED AT ABBOTT DIAGNOSTICS SCARBOROUGH, INC. ON RETAINED KIT LOT 1025742 WITH INTERNAL POSITIVE QUALITY CONTROL SAMPLES AND NEGATIVE QUALITY CONTROL SWABS. ALL TEST RESULTS WERE VALID AND PERFORMED AS EXPECTED. ADDITIONALLY, THE MANUFACTURING RECORDS AND QUALITY CONTROL RELEASE TESTING WAS REVIEWED FOR TEST KIT PART NUMBER 191-000 / LOT: 1025742, TEST BASE PART NUMBER 190-430 / LOT:1025742. THE LOT MET THE REQUIRED RELEASE SPECIFICATIONS. A REVIEW OF THE COMPLAINTS REPORTED AS FALSE POSITIVE PATIENT RESULTS (CONFIRMED AND UNCONFIRMED, CONFLICTING RESULTS) RELATED TO KIT LOT M149779 SHOWED THAT THE COMPLAINT RATE IS (B)(4). ABBOTT DIAGNOSTICS SCARBOROUGH WAS UNABLE TO DETERMINE THE EXACT ROOT CAUSE OF THE REPORTED ISSUE. HOWEVER A POSSIBLE ASSIGNABLE ROOT CAUSE IS SAMPLE INTERFERENCE OR CROSS CONTAMINATION.

Additional Manufacturer Narrative · 1

THE INVESTIGATION IS STILL IN PROGRESS. A SUPPLEMENTAL REPORT WILL BE PROVIDED AFTER COMPLETION.

Description of Event or Problem · 1

THE CUSTOMER REPORTED A FALSE POSITIVE RESULT WITH THE ID NOW COVID-19 ASSAY PERFORMED ON (B)(6) 2021 ON A NASOPHARYNGEAL SWAB (SEC COPAN). CONFIRMATION TESTING WAS PERFORMED ON INGENIUS INSTRUMENT WITH THE NEW SAMPLE GENERATED A NEGATIVE RESULT. THE CUSTOMER STATED THE PATIENT WAS NOT SYMPTOMATIC TO COVID-19 AND HAD SYMPTOMS FOR HYPOTENSION MALAISE. CUSTOMER CONFIRMED THAT PATIENT WAS HOSPITALIZED IN COVID ZONE FOR 24 HOURS ON (B)(6) 2021 FOR CONFIRMATION ASSAY RESULT AND RELEASED ON (B)(6) 2021 IN A STABLE CONDITION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1178708 ID NOW COVID 19 ASSAY MOLECULAR IVD FOR ID NOW COVID-19 QJR ABBOTT DIAGNOSTICS SCARBOROUGH, INC. 1025742 10811877011320

Patients

Seq Age Sex Outcome Treatment
1 89 YR Female Hospitalization