SURDIAL X
Report
- Report Number
- 3016250252-2021-00005
- Event Type
- Malfunction
- Date Received
- August 4, 2021
- Date of Event
- March 25, 2021
- Report Date
- December 2, 2022
- Manufacturer
- SHIBUYA (SBY)
- Product Code
- KDI
- PMA / PMN Number
- K182940
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UK
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
THE MANUFACTURER EVALUATED THE DEVICE IN QUESTION AND DETERMINED THAT THE DEVICE IS SIMILAR TO 510K# K182940, THEREFORE MDR REPORTABILITY WAS DETEMINED ON JULY 8, 2021.
IT WAS REPORTED THAT DURING A DIALYSIS SESSION (ULTRA FILTRATION(UF) GOAL OF 0.5 L OVER 4.5 HOURS) RUN, THE MACHINE WAS SHOWN TO BE REMOVING 3 TIMES AS MUCH FLUID AS PROGRAMMED. AS A CONSEQUENCE, THE PATIENT BECAME UNWELL AFTER HIS BLOOD PRESSURE DROPPED. HE WAS CONSCIOUS, NAUSEATED AND THEN SUBSEQUENTLY VOMITED. THEREFORE, THE MEDICAL STAFF ADMINISTERED APPROXIMATELY 500MLS - 1 LITRE OF SALINE. THE PATIENT THEN RECOVERED AND THE ULTRA FILTRATION WAS STOPPED FOR THE REMAINING TIME ON DIALYSIS. AFTER THE TREATMENT THE PATIENT WAS IN A SATISFACTORY CONDITION TO GO HOME.
THE MANUFACTURER EVALUATED THE DEVICE IN QUESTION AND DETERMINED THAT THE DEVICE IS SIMILAR TO 510K# K182940, THEREFORE MDR REPORTABILITY WAS DETEMINED ON JULY 8, 2021.
IT WAS REPORTED THAT DURING A DIALYSIS SESSION (ULTRA FILTRATION(UF) GOAL OF 0.5 L OVER 4.5 HOURS) RUN, THE MACHINE WAS SHOWN TO BE REMOVING 3 TIMES AS MUCH FLUID AS PROGRAMMED. AS A CONSEQUENCE, THE PATIENT BECAME UNWELL AFTER HIS BLOOD PRESSURE DROPPED. HE WAS CONSCIOUS, NAUSEATED AND THEN SUBSEQUENTLY VOMITED. THEREFORE, THE MEDICAL STAFF ADMINISTERED APPROXIMATELY 500 MLS - 1 LITRE OF SALINE. THE PATIENT THEN RECOVERED AND THE ULTRA FILTRATION WAS STOPPED FOR THE REMAINING TIME ON DIALYSIS. AFTER THE TREATMENT THE PATIENT WAS IN A SATISFACTORY CONDITION TO GO HOME.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1175111 | SURDIAL X | HEMODIALYSIS MACHINE | KDI | SHIBUYA (SBY) | SURDIAL-X2-HD-UJ |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown | Required Intervention |