BLUNT TIP SCREW, 4X46MM
Report
- Report Number
- 0009613350-2021-00378
- Event Type
- Injury
- Date Received
- August 4, 2021
- Report Date
- September 22, 2021
- Manufacturer
- ZIMMER SWITZERLAND MANUFACTURING GMBH
- Product Code
- HSB
- UDI-DI
- 00889024505476
- PMA / PMN Number
- K200814
- Removal / Correction Number
- N/A
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- PHYSICIAN
Narratives
INVESTIGATION RESULTS WERE MADE AVAILABLE. 1. REVIEW OF EVENT DESCRIPTION: IT WAS REPORTED THAT THE PATIENT WAS IMPLANTED WITH ANN NAIL ON (B)(6) 2021. AFTER 2 MONTHS FROM INITIAL SURGERY, THE SURGEON OBSERVED THAT THE 3RD SCREW OF THE PROXIMAL SCREWS WAS BACKED OUT FROM THE IMPLANTED POSITION. THE SURGEON KEEPS MONITORING THE PATIENT CONDITION. NO REVISION SURGERY IS PLANNED SO FAR. HARM: S2 - INSTABILITY, MINOR. HAZARDOUS SITUATION: LOSS OF FIXATION OR NON-INTEGRATION OF AN IMPLANT. 2. REVIEW OF RECEIVED DATA: PATIENT WAS IMPLANTED WITH ANN NAIL ON (B)(6) 2021. AFTER 2 MONTHS FROM INITIAL SURGERY, THE SURGEON OBSERVED THAT THE 3RD SCREW OF THE PROXIMAL SCREWS WAS BACKED OUT FROM THE IMPLANTED POSITION. THE SURGEON KEEPS MONITORING THE PATIENT CONDITION. NO REVISION SURGERY IS PLANNED SO FAR. 3. PRODUCT EVALUATION: NO PRODUCT WAS RETURNED; THEREFORE, VISUAL AND DIMENSIONAL EVALUATION COULD NOT BE PERFORMED. 4. REVIEW OF PRODUCT DOCUMENTATION: DEVICE PURPOSE: THIS DEVICE IS INTENDED FOR TREATMENT. PRODUCT COMPATIBILITY: THE PRODUCT COMBINATION WAS APPROVED BY ZIMMER BIOMET. DHR REVIEW: REVIEW OF THE DEVICE HISTORY RECORDS IDENTIFIED NO DEVIATIONS OR ANOMALIES DURING MANUFACTURING. RAW MATERIAL CERTIFICATE: THE RAW MATERIAL CERTIFICATE WAS REVIEWED WITH NO ANOMALIES NOTED. SURGICAL TECHNIQUE SAP: THE SURGICAL TECHNIQUE 197-GLBL-EN EXPLAINS THAT THE LOCKING OF THE CORELOCK IS DONE USING THE CORELOCK DRIVER WITH TORQUE LIMITING HANDLE. "TURN SLOWLY CLOCKWISE TO TIGHTEN AND ENGAGE THE CORELOCK MECHANISM UNTIL A CLICK IS FELT FROM THE TORQUE LIMITING HANDLE." 5. CONCLUSION: IT WAS REPORTED THAT THE PATIENT WAS IMPLANTED WITH ANN NAIL ON (B)(6) 2021. AFTER 2 MONTHS FROM INITIAL SURGERY, THE SURGEON OBSERVED THAT THE 3RD SCREW OF THE PROXIMAL SCREWS WAS BACKED OUT FROM THE IMPLANTED POSITION. THE SURGEON KEEPS MONITORING THE PATIENT CONDITION. NO REVISION SURGERY IS PLANNED SO FAR. NEITHER X-RAYS, OPERATIVE NOTES, OFFICE VISIT NOTES, NOR DEVICES OR PHOTOS OF THE DEVICES WERE RECEIVED; THEREFORE THE CONDITION OF THE COMPONENTS IS UNKNOWN. PATIENT FACTORS THAT MAY HAVE AFFECTED THE PERFORMANCE OF THE COMPONENTS SUCH AS BONE QUALITY, ACTIVITY LEVEL, TYPE OF ACTIVITY (LOW IMPACT VS. HIGH IMPACT), AND RELEVANT MEDICAL HISTORY ARE UNKNOWN. ADHERENCE TO REHABILITATION PROTOCOL IS UNKNOWN. THE LOCKING OF THE CORELOCK DURING THE INITIAL SURGERY HAS NOT BEEN PERFORMED AS SPECIFIED IN THE SURGICAL TECHNIQUE USING ONLY THE CORELOCK DRIVER WITH TORQUE LIMITING HANDLE. INSTEAD, ADDITIONALLY A NON-TORQUE LIMITING SCREWDRIVER WAS USED. IT REMAINS UNKNOWN WHAT THE POTENTIAL EFFECT OF THIS DEVIATION FROM THE SURGICAL TECHNIQUE COULD BE. BASED ON THE INVESTIGATION THE REPORTED EVENT CANNOT BE CONFIRMED. THE QUALITY RECORDS SHOW THAT ALL SPECIFIED CHARACTERISTICS HAVE MET THE SPECIFICATIONS VALID AT THE TIME OF PRODUCTION. THEREFORE, THE INVESTIGATION DID NOT IDENTIFY A NONCONFORMANCE OR A COMPLAINT OUT OF BOX (COOB). BASED ON THE INVESTIGATION IT COULD BE ASSUMED THAT FURTHER POSSIBLE CONTRIBUTING FACTORS TO THE MIGRATION OF THE SCREW MIGHT BE MULTIFACTORIAL RELATED TO EITHER PATIENT CONDITION AND BEHAVIOUR, IMPLANTATION PROCEDURE OR DESIGN FEATURES. IF AND TO WHAT EXTENT ANY OF THESE ASPECTS MAY HAVE INFLUENCED THE BACKING OUT OF THE SCREW REMAINS UNKNOWN. THE NEED FOR CORRECTIVE MEASURES IS NOT INDICATED FOR THE TIME BEING AND ZIMMER SWITZERLAND MANUFACTURING GMBH CONSIDERS THIS CASE AS CLOSED. ZIMMER BIOMET'S REFERENCE NUMBER OF THIS FILE IS (B)(4).
ADDITIONAL INFORMATION WHICH WAS RECEIVED ON JUL 21, 2021. D10- MEDICAL PRODUCT: TORQUE LIMITING HANDLE; ITEM# : 27923; LOT# : UNKNOWN. LG CANN SCREWDRIVER HANDLE; ITEM# : 214149000; LOT# : UNKNOWN. THIS FOLLOW-UP REPORT IS BEING FILLED TO RELAY ADDITIONAL INFORMATION, WHICH WAS UNKNOWN AT THE TIME OF THE PREVIOUS MEDWATCH. THE MANUFACTURER RECEIVED OTHER SOURCE DOCUMENTS FOR REVIEW. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE AND / OR THE DEVICE(S) BE RETURNED FOR EVALUATION AND AN INVESTIGATION RESULT BE AVAILABLE, THAT CHANGES THIS ASSESSMENT, AN AMENDED MEDICAL DEVICE REPORT WILL BE SUBMITTED. ZIMMER¿S REFERENCE NUMBER OF THIS FILE IS (B)(4).
NO EVENT UPDATE. INVESTIGATION RESULTS ARE NOW AVAILABLE.
PATIENT WAS IMPLANTED WITH ANN NAIL SYSTEM ON AN UNKNOWN SIDE. TWO MONTHS FROM THE INITIAL SURGERY THE SURGEON NOTICED THAT THE THIRD PROXIMAL SCREW HAD BACKED OUT FROM ITS POSITION. THE SURGEON IS MONITORING THE PATIENT'S CONDITION AND A REVISION SURGERY HAS NOT YET BEEN PLANNED.
MEDICAL PRODUCTS: PROXIMAL HUMERUS, RIGHT, LONG, 8.5X220MM; CATALOG#: 47-2496-220-08; LOT#: 3031577; BLUNT TIP SCREW, 4X34MM; CATALOG#: 47-2486-034-40; LOT#: 3010625; BLUNT TIP SCREW, 4X44MM; CATALOG#: 47-2486-044-40; LOT#: 3024701; CORTICAL BONE SCREW, 4X22MM; CATALOG#: 47-2486-122-40; LOT#: 3054244; CORTICAL BONE SCREW, 4X22MM; CATALOG#: 47-2486-122-40; LOT#: 3054448; PROXIMAL HUMERUS NAIL CAP, 0MM; CATALOG#: 47-2488-010-00; LOT#: 3054381. THERAPY DATE: UNKNOWN. THE MANUFACTURER DID NOT RECEIVE X-RAYS OR OTHER SOURCE DOCUMENTS FOR REVIEW. THE MANUFACTURER DID NOT RECEIVE THE DEVICE FOR INVESTIGATION. WHERE LOT NUMBERS WERE RECEIVED FOR THE DEVICES, THE DEVICE HISTORY RECORDS WERE REVIEWED AND FOUND TO BE CONFORMING. A CAUSE FOR THIS SPECIFIC EVENT CANNOT BE ASCERTAINED FROM THE INFORMATION PROVIDED. AS SOON AS SUPPLEMENTAL INFORMATION BECOMES AVAILABLE AN UPDATED REPORT WILL BE SUBMITTED. ZIMMER BIOMET¿S REFERENCE NUMBER OF THIS FILE IS (B)(4).
PATIENT WAS IMPLANTED WITH ANN NAIL SYSTEM ON AN UNKNOWN SIDE. TWO MONTHS FROM THE INITIAL SURGERY THE SURGEON NOTICED THAT THE THIRD PROXIMAL SCREW HAD BACKED OUT FROM ITS POSITION. THE SURGEON IS MONITORING THE PATIENT'S CONDITION AND A REVISION SURGERY HAS NOT YET BEEN PLANNED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1175065 | BLUNT TIP SCREW, 4X46MM | AFFIXUS(R) NATURAL NAIL(R) SYSTEM HUMERAL NAIL | HSB | ZIMMER SWITZERLAND MANUFACTURING GMBH | N/A | 3024713 | 00889024505476 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Hospitalization | SEE H10 NARRATIVE| SEE H10 NARRATIVE. |