PUMP MMT-1780KPK 670G PATHWAY BLACK MG
Report
- Report Number
- 2032227-2021-176433
- Event Type
- Malfunction
- Date Received
- August 4, 2021
- Date of Event
- July 25, 2021
- Report Date
- November 9, 2021
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OZP
- UDI-DI
- 000000763000393786
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- OTHER
- Health Professional
- N
Narratives
CUSTOMER COMPLAINED ON 07/25/2021 INSULIN PUMP ALARMED STUCK BUTTON. DEVICE PASSED SELF TEST AND DISPLACEMENT TEST. INTERMITTENT RESPONSE FROM THE BACK ARROW AND LEFT ARROW BUTTONS DUE TO CORRODED KEYPAD TRACES. DEVICE SUCCESSFULLY DOWNLOADED TO THUS. NO STUCK BUTTON ERROR ALARMS NOTED DURING TESTING. VERIFIED INSULIN PUMP ALARMED STUCK BUTTON ON 07/25/2021 14:27:14.000 IN DEVICE DOWNLOADED HISTORY. DEVICE PASSED THE KEYPAD VOLTAGE TEST. FOUND MOISTURE DAMAGE TO MOTOR ASSEMBLY, ELECTRICAL BOARD 1 AND ELECTRICAL BOARD 2 DURING VISUAL INSPECTION. THE FOLLOWING WERE NOTED DURING VISUAL INSPECTION: CORRODED ELECTRONIC ASSEMBLIES, CORRODED BATTERY TUBE, CORRODED MOTOR HOME SWITCH, SCRATCHED CASE, PILLOWING KEYPAD OVERLAY AND CRACKED CASE (BATTERY TUBE). THE P-CAP/RESERVOIR DOES LOCK PROPERLY. INTERMITTENT RESPONSE FROM THE BACK ARROW AND LEFT ARROW BUTTONS DUE TO CORRODED KEYPAD TRACES. NO STUCK BUTTON ERROR ALARMS NOTED DURING TESTING. VERIFIED PUMP ALARMED STUCK BUTTON ON 07/25/2021 14:27:14.000 IN DEVICE DOWNLOADED HISTORY. FOUND MOISTURE DAMAGE TO MOTOR ASSEMBLY, ELECTRICAL BOARD 1 AND ELECTRICAL BOARD 2 DURING VISUAL INSPECTION. MEDTRONIC, INC. (MEDTRONIC) IS SUBMITTING THIS REPORT TO COMPLY WITH 21 C.F.R. PART 803, THE MEDICAL DEVICE REPORTING REGULATION. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION IN THE TIME ALLOTTED AND HAS PROVIDED AS MUCH INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY "DEFECTS" OR HAS "MALFUNCTIONED". THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA, TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REPORTING PURSUANT TO PART 803. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE IT BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT.
(B)(4). CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD STUCK BUTTON ALARM. CUSTOMER STATED THEY DID NOT PRESS A BUTTON DOWN FOR 3 MINUTES OR LONGER. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1175233 | PUMP MMT-1780KPK 670G PATHWAY BLACK MG | ARTIFICIAL PANCREAS DEVICE SYSTEM, SINGLE HORMONAL CONTROL | OZP | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-1780KPK | HG57ZZQ | 000000763000393786 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 37 YR | Female |