FDA Adverse Event Malfunction Summary report: N

EVA VITRECTOMY PACK VGPC INPUT 25G

MDR report key: 12273500 · Received August 4, 2021

Report

Report Number
1222074-2021-00060
Event Type
Malfunction
Date Received
August 4, 2021
Date of Event
May 21, 2021
Report Date
September 16, 2021
Manufacturer
D.O.R.C. DUTCH OPHTHALMIC RESEARCH CENTER INT.
Product Code
HQC
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TW
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

IN REGARDS TO THIS EVENT, TWO EVA CARTRIDGES WERE RETURNED FOR INVESTIGATION. INVESTIGATION OF THE RETURNED PRODUCTS REVEALED NO ANOMALIES. ALSO NO PROBLEMS OCCURRED DURING TESTING. THEREFORE THE EVENT CANNOT BE ATTRIBUTED TO THE RETURNED CARTRIDGES. POSSIBLY, THE REPORTED EVENT IS ATTRIBUTABLE TO THE PUMP MODULE SUBJECT TO THE EVENT. PLEASE BE INFORMED THAT THE PUMP MODULE IS CURRENTLY BEING INVESTIGATED UNDER (B)(4) / NCA REFERENCE NUMBER 1222074-2021-00048. THE CURRENT COMPLAINT IS CONSIDERED A DUPLICATE OF (B)(4) BUT THIS WAS NOT PREVIOUSLY IDENTIFIED AS DIFFERENT PRODUCTS WERE REFERENCED. THE RISK IDENTIFIED IS INCLUDED IN THE RISK MANAGEMENT DOCUMENTATION. TREND ANALYSIS INDICATES THAT THE PRODUCT IS PERFORMING WITHIN ANTICIPATED RATES. THEREFORE, NO REMEDIAL OR CORRECTIVE/PREVENTIVE ACTIONS WILL BE UNDERTAKEN AT THIS MOMENT. COMPLAINTS WILL BE CLOSELY MONITORED TO IDENTIFY ANY SIGNIFICANT ADVERSE TRENDS.

Description of Event or Problem · 0

IT HAS BEEN REPORTED THAT DURING PROCEDURE THE MACHINE GAVE AN ERROR MESSAGE. THE USER WAS UNABLE TO SOLVE THE ERROR AND IT WAS DECIDED TO ABORT THE SURGERY. IT IS YET UNKNOWN IF GENERAL ANESTHETIC WAS ADMINISTERED. NO PATIENT HARM OCCURRED.

Additional Manufacturer Narrative · 1

THE COMPLAINT IS UNDER INVESTIGATION. IT IS UNKNOWN IF THE PRODUCT WILL BE RETURNED AT THIS POINT.

Description of Event or Problem · 1

IT HAS BEEN REPORTED THAT DURING PROCEDURE THE MACHINE GAVE AN ERROR MESSAGE. THE USER WAS UNABLE TO SOLVE THE ERROR AND IT WAS DECIDED TO ABORT THE SURGERY. IT IS YET UNKNOWN IF GENERAL ANESTHETIC WAS ADMINISTERED. NO PATIENT HARM OCCURRED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1171556 EVA VITRECTOMY PACK VGPC INPUT 25G SURGICAL PACKS AND KITS HQC D.O.R.C. DUTCH OPHTHALMIC RESEARCH CENTER INT. 2000419253

Patients

Seq Age Sex Outcome Treatment
1