FDA Adverse Event
Other
Summary report: N
GE
MDR report key: 1227021
·
Received November 11, 2008
Report
- Report Number
- MW5008950
- Event Type
- Other
- Date Received
- November 11, 2008
- Date of Event
- September 4, 2008
- Report Date
- November 11, 2008
- Manufacturer
- GE CO, DATEX-OHMEDA, INC.
- Product Code
- BSZ
- Product Problem
- Yes
- Report Source
- Voluntary report
- Reporter Location
- MN, US
- Reporter Occupation
- NURSE
Narratives
Description of Event or Problem · 1
ANESTHESIA MACHINE SHUTDOWN IN THE MIDDLE OF A CASE. COMPANY TECHNICIAN CAME OUT, AND REPLACED SWITCH. COMPANY TECHNICIAN CAME OUT, STANDBY SWITCH SUSPECTED TO BE CAUSE OF SHUTDOWN AFTER LOGS REVIEWED WITH TECH. THE STANDBY SWITCH WAS REPLACED. TESTED TO MFR'S PROCEDURE FOR AISYS, ALL PASSED. DESCRIPTION: ASSY-MSN, ASSY-MSN, SWITCH D-O ON/STANDBY ASSY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | GE | ANESTHESIA MACHINE | BSZ | GE CO, DATEX-OHMEDA, INC. | 1011-9000-000 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | * | Other |