LM-900 N2O CE CRYOSURG SY
Report
- Report Number
- 1216677-2021-00155
- Event Type
- Malfunction
- Date Received
- August 3, 2021
- Date of Event
- June 14, 2021
- Report Date
- January 30, 2023
- Manufacturer
- COOPERSURGICAL, INC.
- Product Code
- GEH
- UDI-DI
- 00888937009965
- PMA / PMN Number
- K854767
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NU
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
INVESTIGATION: REVIEW DHR, INSPECT RETURNED SAMPLES. DISTRIBUTION HISTORY: THIS COMPLAINT UNIT WAS MANUFACTURED AT CSI ON 1/23/2019 UNDER WO #(B)(4) AND SHIPPED ON 7/30/2020. MANUFACTURING RECORD REVIEW: DHR 241713 WAS REVIEWED AND NO NON-CONFORMITIES, RELATED TO THE COMPLAINT CONDITION, WERE NOTED. INCOMING INSPECTION REVIEW: NOT APPLICABLE. SERVICE HISTORY RECORD: NO ADDITIONAL SERVICE HISTORY RECORDS FOUND FOR THIS UNIT. HISTORICAL COMPLAINT REVIEW: A REVIEW OF THE 2-YEAR COMPLAINT HISTORY SHOWED SIMILAR REPORTED COMPLAINT CONDITION. PRODUCT RECEIPT: THE COMPLAINT UNIT WAS RETURNED ON A REPAIR. VISUAL EVALUATION: VISUAL EXAMINATION OF THE COMPLAINT UNIT REVEALED NO OUTER PHYSICAL DAMAGE. UPON OPENING THE UNIT THE TUBING WAS FOUND DAMAGED FROM RUPTURING. FUNCTIONAL EVALUATION: COMPLAINT UNIT WAS FUNCTIONALLY EVALUATED AND FOUND NOT TO FUNCTION PROPERLY. ROOT CAUSE: THE ASSEMBLY OF THE LINES INSIDE THE HOUSING REQUIRES THE NYLON TUBING TO BE TURNED AND CURVED TO MATE WITH THE FITTINGS. THE LINE MAY HAVE BEGUN TO KINK AND EVENTUALLY PREVENTED NORMAL GAS FLOW OR BLOCKAGE CREATING A HIGH PRESSURE POINT. HOWEVER, A ROOT CAUSE IS NOT DEFINITIVELY DETERMINED. CORRECTIVE ACTIONS THE UNIT WAS REPAIRED, TESTED TO SPECIFICATIONS AND RETURNED TO THE CUSTOMER UNDER WARRANTY. COOPERSURGICAL WILL CONTINUE TO MONITOR THIS COMPLAINT CONDITION FOR TRENDS. NO FURTHER TRAINING REQUIRED. WAS THE COMPLAINT CONFIRMED? YES.
CUSTOMER STATED: "GAS LEAK IN METER". LM-900 N20 CE CRYOSURG SY (B)(4).
COOPERSURGICAL INC. IS CURRENTLY INVESTIGATING THE REPORTED CONDITION.
CUSTOMER STATED "GAS LEAK IN METER". REPAIR TECH STATED "FOUND - HP LINE FORM ON/OFF TO PRESSURE GAUGE BLOWN OUT - REPLACED HP LINE". (B)(4). GAS EXPOSURE DUE TO OVER PRESSURE OF SUPPLY TANK RELATES TO A RISK OF 6 PRE AND 11 POST INDICATING A REPORTABLE EVENT PER TECH-035. 1216677-2021-00155 LM-900 N2O CE CRYOSURG SY 50501 (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1169265 | LM-900 N2O CE CRYOSURG SY | LM-900 N2O CE CRYOSURG SY | GEH | COOPERSURGICAL, INC. | 50501 | N/A | 00888937009965 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown | Other |