FDA Adverse Event Injury Summary report: N

LINER: CC E CC LIGHT 01.26.2844HCT FLAT PE HC LINER 28 / E

MDR report key: 12267572 · Received August 3, 2021

Report

Report Number
3005180920-2021-00651
Event Type
Injury
Date Received
August 3, 2021
Date of Event
July 20, 2021
Report Date
October 11, 2021
Manufacturer
MEDACTA INTERNATIONAL SA
Product Code
LZO
UDI-DI
07630030807176
PMA / PMN Number
K103352
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
SZ
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

CLINICAL EVALUATION PERFORMED BY MEDACTA MEDICAL AFFAIRS MANAGER ON 14 SEPTEMBER 2021: HIP DISLOCATION 1 YEAR AND 5 MONTHS AFTER PRIMARY IMPLANTATION. HIP DISLOCATION MAY BE DUE TO COMPONENT POSITIONING OR INSUFFICIENT SOFT TISSUE TENSIONING; IT CAN HARDLY BE CAUSED BY STANDARD DEVICES.

Additional Manufacturer Narrative · 1

BATCH REVIEW PERFORMED ON 20.07.2021. LOT 181142: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 29-MAY-2018. EXPIRATION DATE: 2023-05-14. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN SOLD WITHOUT SIMILAR REPORTED EVENT. OTHER DEVICE INVOLVED: BATCH REVIEW PERFORMED ON 20.07.2021. BALL HEADS: MECTACER 01.29.202 BIOLOX DELTA CERAMIC BALL HEAD 12/14 Ø 28 SIZE M 0 (K112115) LOT. 1901932: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 17-JUN-2019. EXPIRATION DATE: 2024-05-27. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN SOLD WITHOUT SIMILAR REPORTED EVENT.

Description of Event or Problem · 1

REVISION SURGERY PLANNED 1 YEAR AND 5 MONTHS AFTER THE PRIMARY DUE TO DISLOCATION. LINER AND HEAD HAS BEEN REVISED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1162566 LINER: CC E CC LIGHT 01.26.2844HCT FLAT PE HC LINER 28 / E HIP LINER FLAT LZO MEDACTA INTERNATIONAL SA 01.26.2844HCT 181142 07630030807176

Patients

Seq Age Sex Outcome Treatment
1 73 YR Required Intervention