FDA Adverse Event Malfunction Summary report: N

XTRA AUTOTRANSFUSION SYSTEM

MDR report key: 12259322 · Received August 2, 2021

Report

Report Number
9680841-2021-00021
Event Type
Malfunction
Date Received
August 2, 2021
Date of Event
June 24, 2021
Report Date
September 30, 2021
Manufacturer
SORIN GROUP ITALIA SRL
Product Code
CAC
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

LIVANOVA RECEIVED REPORT STATING THAT, WHEN INSERTING THE TERMINAL NEEDLE OF WASH LINE TO THE SALINE BAG FOR SETUP OF BOWL SET, OPERATOR CUT HIS FINGER WITH THE EDGE OF THE CORNER PART. VISUAL INSPECTION OF THE RETURNED SPIKE NEEDLES FOUND NEITHER ANY MATERIAL INTEGRITY PROBLEM NOR ANY CUTTING EDGES (NO SHARP POINT NOR BURR). VERIFICATION OF MANUFACTURING RECORDS CONFIRMED THAT CLAIMED DEVICE WAS RELEASED CONFORMING TO PRODUCT SPECIFICATIONS. ANALYSIS OF COMPLAINTS DATABASE REVEALED NO OTHER SIMILAR EVENT RELEVANT TO THE COMPLAINED LOT. BASED ON INVESTIGATION RESULTS, NO DIRECT RELATIONSHIP BETWEEN USER HURT AND ANY MATERIAL NON-CONFORMITY WAS ESTABLISHED. A POSSIBLE ROOT CAUSE OF REPORTED EVENT WAS A ROUGH HANDLING BY OPERATOR. SINCE NO DEVICE MALFUNCTION COULD BE CONFIRMED, NO ACTION WAS DEEMED NECESSARY. LIVANOVA WILL MAINTAIN MONITORING THE MARKET.

Description of Event or Problem · 0

SEE INTIAL REPORT.

Additional Manufacturer Narrative · 1

PATIENT INFORMATION WERE NOT PROVIDED. THE PRODUCT ITEM 04255J - PROCEDURE SET TX/125 IS NOT DISTRIBUTED IN THE USA AND IT IS SIMILAR ITEM 04255. THE EXPIRY DATE REFERS TO THE STERILE FINISHED PRODUCT. THE COMPLAINED PRODUCT ITEM 04255J - PROCEDURE SET TX/125 IS NOT DISTRIBUTED IN THE USA, THEREFORE THE UDI IS NOT APPLICABLE. THE PRODUCT ITEM 04255J - PROCEDURE SET TX/125 IS SIMILAR TO THE 04255, WHICH IS DISTRIBUTED IN THE USA, FOR WHICH THE DEVICE IDENTIFIER IS (B)(4). THE PRODUCT ITEM 04255J - PROCEDURE SET TX/125 IS NOT DISTRIBUTED IN THE USA, BUT IT IS SIMILAR TO THE ITEM 04255 - PROCEDURE SET TX/125, WHICH IS DISTRIBUTED IN THE USA (510(K) NUMBER: K101586). THE DEVICE MANUFACTURE DATE REFERS TO MANUFACTURE DATE OF THE STERILE, FINISHED PROCEDURE SET TX/125. SORIN GROUP (B)(4) MANUFACTURES THE PROCEDURE SET TX/125. THE INCIDENT OCCURRED IN (B)(6). THE DEVICE HAS BEEN REQUESTED FOR INVESTIGATION AND NOT YET RECEIVED. IF ANY ADDITIONAL INFORMATION PERTINENT TO THE REPORTED EVENT IS OBTAINED, IT WILL BE PROVIDED IN A SUPPLEMENTAL REPORT.

Description of Event or Problem · 1

SORIN GROUP (B)(4) HAS RECEIVED A REPORT THAT, DURING SETTING THE WASH SET PROCEDURE SET TX/125, WHILE ATTEMPTING TO SPIKE THE WASHING LINE INTO THE SALINE PACK, THE GLOVE AND FINGER-SKIN OF THE HEALTH OPERATOR GOT HURT WITH THE CORNERS. THE CUSTOMER FELT THAT THE CORNER PART WAS SHARPER THAN IN A OTHER PRODUCT OF THE SAME CATALOGUE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1160634 XTRA AUTOTRANSFUSION SYSTEM APPARATUS, AUTOTRANSFUSION CAC SORIN GROUP ITALIA SRL 2010290094

Patients

Seq Age Sex Outcome Treatment
1