FDA Adverse Event Injury Summary report: N

CUP: MPACT ACETABULAR SHELL 50 TWO-HOLES

MDR report key: 12258956 · Received August 2, 2021

Report

Report Number
3005180920-2021-00658
Event Type
Injury
Date Received
August 2, 2021
Date of Event
July 19, 2021
Report Date
August 2, 2021
Manufacturer
MEDACTA INTERNATIONAL SA
Product Code
LPH
UDI-DI
07630030810763
PMA / PMN Number
K132879
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
IS
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 1

BATCH REVIEW PERFORMED ON 28 JULY 2021: LOT 2010311: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 26-JAN-2021. EXPIRATION DATE: 2026-01-12. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN SOLD WITHOUT ANY SIMILAR REPORTED EVENT. OTHER DEVICES INVOLVED: BATCH REVIEW PERFORMED ON 28 JULY 2021: LINER: MPACT 01.32.3241HCT FLAT PE HC LINER Ø32/D (K103721) LOT 2100154: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 17-FEB-2021. EXPIRATION DATE: 2026-02-01. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN SOLD WITHOUT ANY SIMILAR REPORTED EVENT. SCREWS: MPACT 01.32.6535 CANCELLOUS BONE SCREW, FLAT HEAD Ø 6,5 L 35 (K103721) LOT 2013780: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 12-FEB-2021. EXPIRATION DATE: 2026-02-01. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN SOLD WITHOUT ANY SIMILAR REPORTED EVENT. STEM: QUADRA -R 01.12.065 CEMENTLESS, HA COATED REVISION STEM SIZE 5 (K082792) LOT 1810350: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 27-FEB-2019. EXPIRATION DATE: 2024-02-18. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN SOLD WITHOUT ANY SIMILAR REPORTED EVENT.

Description of Event or Problem · 1

REVISION SURGERY 1 MONTH AFTER PRIMARY DUE TO INFECTION. STAPH EPIDERMIS WAS THE PATHOGEN. THE SURGERY WAS COMPLETED SUCCESSFULLY, ALL IMPLANT REVISED SUCCESSFULLY AND TIBIAL SPACER IMPLANTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1160884 CUP: MPACT ACETABULAR SHELL 50 TWO-HOLES ACETABULAR SHELL LPH MEDACTA INTERNATIONAL SA 01.32.150DH 2010311 07630030810763

Patients

Seq Age Sex Outcome Treatment
1 74 YR Required Intervention