FDA Adverse Event Malfunction Summary report: N

BD PLASTIPAK 50ML CONCENTRIC LUER LOCK SYRINGE

MDR report key: 12256532 · Received July 30, 2021

Report

Report Number
3003152976-2021-00441
Event Type
Malfunction
Date Received
July 30, 2021
Date of Event
June 23, 2021
Report Date
August 20, 2021
Manufacturer
BECTON DICKINSON, S.A.
Product Code
FMF
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

H6: INVESTIGATION SUMMARY NO PHOTOS OR PHYSICAL SAMPLES THAT DISPLAY THE REPORTED CONDITION WERE AVAILABLE FOR INVESTIGATION. A DEVICE HISTORY REVIEW WAS PERFORMED FOR THE REPORTED 2011035, NO DEVIATIONS OR NON-CONFORMANCES WERE IDENTIFIED DURING THE MANUFACTURING PROCESS THAT COULD HAVE CONTRIBUTED TO THE REPORTED ISSUE. TEN RETAINED SAMPLES OF LOT 2011035 WERE USED FOR ADDITIONAL EVALUATION. THE PRODUCT WAS VISUALLY INSPECTED, NO DEFECTS OR DAMAGE WAS NOTED, THE STOPPER WAS PROPERLY ASSEMBLED ONTO THE PLUNGER ROD, AND NO LEAK OCCURRED DURING FUNCTIONAL TESTING. FINAL PRODUCTS IN THIS MANUFACTURING LINE, FOR THIS REFERENCE ARE SAMPLED AND THEY ARE SUBJECTED TO VISUAL AND FUNCTIONAL INSPECTIONS DURING THE DIFFERENT MANUFACTURING SUB-PROCESSES ACCORDING TO PROCEDURES. BASED ON THE AVAILABLE INFORMATION WE ARE NOT ABLE TO DETERMINE A ROOT CAUSE AT THIS TIME. COMPLAINTS RECEIVED FOR THIS DEFECT AND DEVICE WILL BE MONITORED BY OUR QUALITY TEAM FOR SIGNS OF EMERGING TRENDS.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE BD PLASTIPAK¿ 50ML CONCENTRIC LUER LOCK SYRINGE EXPERIENCED A LOOSE PLUNGER, AND LEAKAGE PAST THE PLUNGER. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: THE PLUNGER DOES NOT FULFILL ITS SEALING ROLE. THE CYTOTOXIC PRODUCT HAS LEAKED ON THE WORK SURFACE. LOSS OF PRODUCT. NO PATIENT IMPACT BUT RISK OF EXPOSURE FOR STAFF.

Additional Manufacturer Narrative · 1

A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE BD PLASTIPAK¿ 50ML CONCENTRIC LUER LOCK SYRINGE EXPERIENCED A LOOSE PLUNGER, AND LEAKAGE PAST THE PLUNGER. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: THE PLUNGER DOES NOT FULFILL ITS SEALING ROLE. THE CYTOTOXIC PRODUCT HAS LEAKED ON THE WORK SURFACE. LOSS OF PRODUCT. NO PATIENT IMPACT BUT RISK OF EXPOSURE FOR STAFF.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1148891 BD PLASTIPAK 50ML CONCENTRIC LUER LOCK SYRINGE SYRINGE FMF BECTON DICKINSON, S.A. 2011035

Patients

Seq Age Sex Outcome Treatment
1